Table 4.
Main reasons for rejecting validation studies by STRIDE BP (129 studies) and for rejecting studies that were published in PubMed journals as “Pass” (79 studies)
Main reasons of validation study rejection by STRIDE BP* | STRIDE BP rejected studies (%) | ||
---|---|---|---|
All (N = 129) |
Published as “Pass” (N = 79) |
||
Device inaccurate | Key validation criteria not fulfilled | 57 (44%) | 7 (9%) |
Incorrect study execution | Inappropriate sample size | 16 (12%) | 11 (14%) |
Incorrect study execution | Single observer | 22 (17%) | 18 (23%) |
Incorrect study execution | BP measurements protocol violation (sequence, number, repeats) | 37 (29%) | 29 (37%) |
Incorrect study execution | Inadequate reference device, or cuffs for reference and test device | 13 (10%) | 8 (10%) |
Incorrect data analysis | Inappropriate data analysis | 27 (21%) | 27 (34%) |
Incorrect data reporting | Key validation criteria not reported | 35 (27%) | 30 (38%) |
Some studies with more than one major reason for rejection.