. 2019 Oct 26;21(11):1616–1622. doi: 10.1111/jch.13710

Table 4.

Main reasons for rejecting validation studies by STRIDE BP (129 studies) and for rejecting studies that were published in PubMed journals as “Pass” (79 studies)

Main reasons of validation study rejection by STRIDE BP* STRIDE BP rejected studies (%)

All

(N = 129)

Published as “Pass”

(N = 79)

Device inaccurate Key validation criteria not fulfilled 57 (44%) 7 (9%)
Incorrect study execution Inappropriate sample size 16 (12%) 11 (14%)
Incorrect study execution Single observer 22 (17%) 18 (23%)
Incorrect study execution BP measurements protocol violation (sequence, number, repeats) 37 (29%) 29 (37%)
Incorrect study execution Inadequate reference device, or cuffs for reference and test device 13 (10%) 8 (10%)
Incorrect data analysis Inappropriate data analysis 27 (21%) 27 (34%)
Incorrect data reporting Key validation criteria not reported 35 (27%) 30 (38%)
*

Some studies with more than one major reason for rejection.