Table 1.

Types of office blood pressure measurement

Sources of error I. Auscultatory clinical practice II. Automated attended clinical practice III. Auscultatory or automated research setting IV. Automated unattended clinical practice
Standardization + +++ +++
Resting period Inconsistent Inconsistent Yes Yes
Devices maintenance and accuracy Variable Usually ok Good Usually ok
Cuff sizes 1‐2 1‐2 3+ 2+
Talking during resting and between readings Common Common No No
Observer error, bias, terminal digit preference Yes No No No
Number of readings 1‐2 1‐3 2‐3 3‐6
Hypertension threshold (mmHg) Significantly higher than 140/90 Equal or higher than 140/90 Equal or lower than 140/90 Equal or lower than 135/85
Time required 1′‐5′ 1′‐5′ ~10′ ~10′
Applicability for general practice Yes Yes Not possible Not for all practices (requires device, office space, time)
Outcome data No Strong Strong Few