Performance specifications of the two automated assays for serum C-terminal telopeptide of type I collagen (CTX) [references 24,27,28].

β-CTX assay Measuring range Lower limit of detection Intra-assay CV Inter-assay CV
iSYS, IDS, UK 50–6000 ng/L 20 ng/L 2.7–3.7% 2.5–5.2%
Roche Diagnostics, Germany 10–6000 ng/L 10 ng/L 1.2–4.1% <5.7%
CrossLaps®, IDS, UK 20–3380 ng/mL 20 ng/L <2.5% 2.2–5.5%