Table V.
IAPV parameters; we suggest starting: Pbelt 0-70 Hpa (at the beginning 30-40 Hpa); select desired Ti (during the Ti setted, the PBAir will be deflated, while the patient will be able to inhale); back up rate as desidered; rise time usually 1.0s; Expiratory time (abdominal compression) will be linked to the back up rate and inspiratory time setted. For example: setted inspiratory time 1.5 sec, Fr 15 bpm, derivative expiratory time will be 2.5 sec.
Intermittent abdominal pressure ventilator (LunaBelt) | ||
---|---|---|
Mode | Timed | Spontaneous/timed |
Pression belt | 0-70 hPa | 0-70 hPa |
Time inspiratory | 0.3-5.0 sec | na |
Time inspiratory minimum | na | 0.3-3.0 sec |
Time inspiratory maximum | [(60/Freq) - 0.6 sec] | [(60/Freq - 0.6 sec)] |
Time espiratory minimum | na | 0-1.5 sec |
Back-up Frequency | 1-60 bpm | 1-60 bpm |
Frequency maximum | [60/(Tinsp + 0.6 sec)] | [60/(Tinsp + 0,6 sec) |
Rise time | 0.1-1.0 sec | 0.1-1.0 sec |
Trigger inspiratory (nasal cannula) | na | Auto |
Trigger espiratory (nasal cannula) | na | Auto |