IAPV parameters; we suggest starting: Pbelt 0-70 Hpa (at the beginning 30-40 Hpa); select desired Ti (during the Ti setted, the PBAir will be deflated, while the patient will be able to inhale); back up rate as desidered; rise time usually 1.0s; Expiratory time (abdominal compression) will be linked to the back up rate and inspiratory time setted. For example: setted inspiratory time 1.5 sec, Fr 15 bpm, derivative expiratory time will be 2.5 sec.

Intermittent abdominal pressure ventilator (LunaBelt)
Mode Timed Spontaneous/timed
Pression belt 0-70 hPa 0-70 hPa
Time inspiratory 0.3-5.0 sec na
Time inspiratory minimum na 0.3-3.0 sec
Time inspiratory maximum [(60/Freq) - 0.6 sec] [(60/Freq - 0.6 sec)]
Time espiratory minimum na 0-1.5 sec
Back-up Frequency 1-60 bpm 1-60 bpm
Frequency maximum [60/(Tinsp + 0.6 sec)] [60/(Tinsp + 0,6 sec)
Rise time 0.1-1.0 sec 0.1-1.0 sec
Trigger inspiratory (nasal cannula) na Auto
Trigger espiratory (nasal cannula) na Auto