Table VI.

NPV parameters; we suggest start: Inspiratory pressure of -20, Expiratory pressure from 0 to 5, I:E Ratio from 1:1 to 1:2, back up frequency: set frequency at 2-4 breaths above patient’s own spontaneous rate.

Negative pressure ventilation (Negavent)
Mode T (timed) ST (spontaneous/timed) Syncro (syncrhronized)
Inspiratory pressure Da -5 a -90 hPa Da -5 a -90 hPa Da -5 a -90 hPa
Expiratory pressure Da +25 a -25 hPa Da +25 a -25 hPa Da +25 a -25 hPa
Back-up frequency 5-60 bpm 5-60 bpm 5-60 bpm
I/E ratio 1.0:9.9-9.9:1.0 1.0:9.9-9.9:1.0 1.0:9.9-9.9:1.0
Trigger inspiratory (nasal cannula) na 1;9 1;9
Trigger espiratory (nasal cannula) na 1;9 na