Table 4.
Ref. | Technique | Substrate Material * | Freq (GHz) | Size (mm) | Tested Material (MUT) | Remark |
---|---|---|---|---|---|---|
[68] | ENGResonator | FR-4 = 4.4 h = 1.6 mm | 2.074 | 40 × 20 | Water-glucose | The concentration range of the glucose is 20–100 mg/mL. |
[67] | RF patch biosensor | N/A | 17.25 and 21.09 | 3.3 × 2 | DI water, D-glucose powder | The concentrations range of the glucose is 50–600 mg/dL. The sensitivity achieved is up to 1.13 MHz and 1.97 MHz per mg/dL, and the detection limits are 26.54 mg/dL and 15.22 mg/dL. |
[70] | Broadband dielectric spectroscopy | N/A | 2 to 18 | N/A | N/A | The concentration is varied from 0 to 10 mg/mL. The sensitivity achieved is 0.15 dB/(mg/mL) and 0.5°/(mg/mL) at a room temperature of 25 °C. |
[69] | Antenna-coupled SRR | N/A | 1.9 | 8.3 | N/A | The tested material is blood plasma with different glucose concentrations. |
[71] | Antenna-driven ring | FR4 t = 35 um | 2 | 30 × 30 | DI water, glucose, NaCl solutions | A 17.5 MHz shifting of the resonance frequency is obtained with a high rate of error of 7.3%. The sensitivity achieved is 0.107 MHz/mg dL. |
[72] | Transmission/reflection line method | N/A | 1 to 1.20 | N/A | N/A | The tested range of water-glucose concentration is from 50 mg/dL to 1000 mg/dL with a frequency range from 100 MHz to 4 GHz. |
[66] | Planar MRR | FR4 Epoxy = 4.4 | 1 | 108.75 × 65 | Glucose percentage | The concentration of water-glucose is tested, and there is a slightly frequency shift. However, the sample is directly tested on the sensor, which increases the measurement error. |
[65] | RMLresonator | N/A | 9.20 | 2 × 0.854 | Glucose percentage | The sensitivity achieved is 1.08 MHz per 1 mg dL. The detection limit is 8.01 mg dL with a quantisation limit of 24.30 mg dL. |
* h = thickness, t = copper thickness, = dielectric constant, and = loss tangent. N/A = Not Available.