Table 1.

Full breast axial 3T protocol, 16-channel breast array.

Sequences 1 2 3 4 5
Series descriptor 3D axial T2 3D axial T1 3D axial DISCO 2D axial DWI 2D axial MDD DWI
Sequence type Fast-Spin Echo Gradient-Echo VIBRANT Gradient-Echo Spin-Echo EPI Spin-Echo EPI
Field of view (cm) 34–38 34–38 34–38 34–38 34–38
Slice thickness (mm) 3 1.1 1.1 3.9 3.9
Gap (mm) 3 0 0 0 2.7
Fat saturation Yes No Yes Yes Yes
TE (ms) Minimum Minimum Minimum Minimum 98
TR (ms) 2500–6000 4–4.5 4–4.5 2000–6000 2700–5000
Matrix size (mm) 512 × 512 512 × 512 512 × 512 256 × 256 128 × 128
Flip angle 111 10 12 90 90
Direction ALL ALL
Number of directions 3 43
b-values (s/mm2) 0, 800 100, 700, 1400, 2000
Frequency direction A/P A/P A/P A/P A/P
Scan time (min:sec) 2:32 1:30 5–6 1 4:02 4:12

Abbreviations: A/P, anterior/posterior; DISCO, differential subsampling with Cartesian ordering; EPI, echo planar imaging; MDD, multidimensional diffusion; VIBRANT, volume imaging for breast assessment. 1 For a total of 10–13 phases with a temporal resolution of 12 ms.