Skip to main content
. 2021 Apr 4;26(7):2073. doi: 10.3390/molecules26072073

Table 2.

Sensitivity and precision of the experimental procedure for analysis by headspace-stir bar sorptive extraction-gas chromatography/mass spectrometry.

Standard LOD
(µg Standard/20 mL vial)
LOQ
(µg Standard/20 mL vial)
Repeatability (% RSD) Reproducibility (% RSD)
Safranal 3.992 × 10−4 1.210 × 10−3 2.338 4.050
Isophorone 1.628 × 10−3 4.935 × 10−3 1.861 3.223
Linalool 2.590 × 10−2 7.847 × 10−2 2.200 3.811
tMCHdO 1 2.650 × 10−3 8.029 × 10−3 1.723 2.984

1 tMCHdO: 2,6,6-trimethyl-2-cyclohexene-1,4-dione. LOD: limit of detection; LOQ: limit of quantification. RSD: relative standard deviation.