Table 2.

GCMS and LCMS parameters.

Gas Chromatography (GC)-Mass Spectrometry (MS) Parameters (Myo-Inositol and NAA)
Liquid Chromatography (LC)- Mass Spectrometry (MS) Parameters (Choline and Creatine)
GAS chromatography (GC) Liquid chromatography (LC)
GC Column

Injection Temp
Constant Gas Flow
Injection Mode
Split Ratio
Injection Volume
Run Time
J&W DB-5MSCapillary 30 m × 250 μm × 0.25 μm
280 °C
1 ml/min
Split
10:1
1 μL
13.3 min
LC Column

LC Column Temp
Injection Tray Temp
Injection Volume
Flow Rate
Run Time
Agilent Zorbax Eclipse C18 Rapid Resolution HD 100 × 2.1 mm 1.8 μ
40 °C
10 °C
5 μL
200 μL
6 min
GC oven Temp Program
temperature 100 °C/min and hold time 1.0 min, oven ramp: 30 °C/min, temperature 320 °C/min, hold time; 5 min, run time; 13.3 min
Retention Times
Myo-Inositol
N-Acetylaspartic Acid
Myo-Inositol-D6
Malic Acid-D3

7.2 min
5.7 min
7.2 min
4.6 min
Retention Times Choline
Creatine
Acetylcholine-D4

1.2 min
1.3 min
1.3 min



Mass spectometry (MS) Mass spectometry (MS)
Source Temperature
Quad Temperature
Acquisition time
230 °C
150 °C
5–13.3 min
Ionisation
Polarity
Spray Voltage Capillary Temp
Auxiliary Gas Flow Sheath Gas Flow
Acquisition time
Scan type:
Scan Width:
Scan Time
Data type
Numbers of Microscans
Collision Gas Pressure:
Divert Valve:
Electrospray (ESI)
Positive
3500 V
350 °C
0 Arb
50 Arb
6 min
SRM
0.200 m/z
0.150 s
Profile

1

1.5 mTorr
into Waste: 0–0.3 min, into MS: 3–6 min