Table 3.

Agreement between assays in classifying blood samples according to analytic benchmarks.

Analytic benchmarks: Roche Elecsys n (%) Abbott Architect n (%) Siemens Vista n (%) P values*
Roche vs. Abbott Roche vs. Siemens Abbott vs. Siemens Overall Comparison
< LOD 578 (56.3) 107 (10.4) 423 (41.2) <0.001 <0.001 <0.001 <0.001
LOD - 99th %tile 375 (36.5) 858 (83.5) 540 (52.6) <0.001 <0.001 <0.001 <0.001
> 99th %tile 74 (7.2) 62 (6.0) 64 (6.2) 0.064 0.157 0.670 0.114
Total 1,027 (100.0) 1,027 (100.0) 1,027 (100.0) N/A N/A N/A N/A
*

Indicating the differences between the assays. LOD=Limit of detection; %tile=Percentile.