Skip to main content
. 2021 Feb 24;62:100053. doi: 10.1016/j.jlr.2021.100053

Table 4.

Performance of three direct LDL-C assays with detection of cholesterol on purified Lp(a) and Lp(a) spiked-in plasma

Input
% Pure Lp(a)-C detected % Spiked-in Lp(a)-C detected
Plasma sample Purified Lp(a) Plasma + purified Lp(a)
TC 155.6 (3.2) 54.9 (1.3) 200.5 (0.6) 100 100
Roche direct LDL-C 97.7 (2.2) 44.4 (0.6) 144.0 (1.1) 87 (3.1) 84 (2.3)
Sekisui direct LDL-C 113.1 (6.4) 57.3 (1.9) 166.8 (3.3) 104 (5.4) 98 (0.2)
Wako direct LDL-C 105.1 (5.2) 49.7 (2.3) 158.9 (2.9) 90 (6.2) 98 (0.1)

Purified Lp(a) with a cholesterol content of 54.9 mg/dl was used for these experiments. % pure Lp(a)-C detected = Lp(a)-C measured by the respective direct LDL-C assay divided by Lp(a)-C measured by the TC assay. % spiked-in Lp(a)-C = cholesterol measured in plasma spiked with pure Lp(a) − cholesterol measured in plasma without added Lp(a) divided by cholesterol measured in purified Lp(a), using each respective assay. Data described are from three separate experiments and expressed as mean (SD). Units are in milligrams per deciliter unless otherwise specified.