Within-run and within-lab precision of the citrate assay.



Citrate (mg/dL)
Within-Run Precisiona Within-Laboratoryb
Low Mean 2.1 2.1
SD 0.2 0.2
%CV 9.3 9.6
High Mean 3.3 3.2
SD 0.2 0.2
%CV 5.8 5.2

a Serum pools tested in 1 run of 20 replicates. b Serum pools tested in 2 runs of duplicates per day for 20 days (n ​= ​80 per analyte). %CV, coefficient of variation expressed as percent.