Table 1 –

Pulse sequence parameters

QISS MRA PDIP-SOS

1.5T 3T 1.5T 3T
Imaging type 2D 2D 3D 3D
Field of view (mm) 400 × 400 400 × 400 410 × 410 416 × 416
Slice thickness (mm) 3 3 0.5 0.5
In-plane resolution (mm) 1.0 × 1.0 1.0 × 1.0 1.0 × 1.0 1.1 × 1.1
Echo time (ms) 1.4 1.7 4.7 2.5
Repetition time (ms) 3.5 3.8 9.6 5.0
Flip angle 90° 90° 4.5° 2.5°
Bandwidth (Hz/pixel) 658 960 300 750
Inversion time (ms) 350 350
GRAPPA 2 2
Readout Cartesian Cartesian Stack-of-Stars Stack-of-Stars
Acquisition time (min) 12* 12* 6.2 5.9

GRAPPA, generalized autocalibrating partially parallel acquisition acceleration factor

Stack-of-stars denotes radial acquisition within each slice-encoding step, and Cartesian sampling in the slice-encoding direction

*

Considering 10 stations, a heart rate of 70 beats per minute, and 15 seconds breath-holds in the 3 abdominal stations