Skip to main content
. Author manuscript; available in PMC: 2022 May 1.
Published in final edited form as: Eur Radiol. 2020 Oct 17;31(5):2778–2787. doi: 10.1007/s00330-020-07386-4

Table 3 –

Subjective image quality ratings and inter-reader agreement

CTA QISS MRA p-value NCCT PDIP-SOS p-value
Rating 4.0
[3.0–4.0]
3.0
[3.0–4.0]
0.0369 4.0
[4.0–4.0]
4.0
[3.0–4.0]
0.0093
κ 0.819
[0.778–0.860]
0.669
[0.605–0.733]
0.891
[0.846–0.924]
0.517
[0.401–0.632]

CTA, Computed tomography angiography; QISS MRA, Quiescent interval slice-selective magnetic resonance angiography; NCCT, non-contrast CT; PDIP-SOS, proton density-weighted in-phase stack of stars MR

Results are reported with [95% confidence interval]