Skip to main content
. Author manuscript; available in PMC: 2022 May 1.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2021 Feb 27;1171:122607. doi: 10.1016/j.jchromb.2021.122607

Table 2.

Dynamic range, linearity, LOD, and precision for vitamin E acetate in synthetic BAL fluid.

Vitamin E acetate
LOD Dynamic Range Linearity Precision (CV%; n = 20)
(ng/mL) (ng/mL) (R2; n = 7) Repeatability Intermediate Precision
25 ng/mL 250 ng/mL 25 ng/mL 250 ng/mL
1.10 10–1000 0.999 2.31 3.28 3.55 8.92