Table 2.

Precision of the parameters are compared using coefficient of variation (CV), reported as percentage and the normalized RMSE (nRMSE). Each row reports on model fitting results obtained from data before applying any correction (no correction) and after applying both distortion and motion correction (Di+MoCo). The coefficient of variation for all parameters and residuals were smallest after applying Di+MoCo.

Model param CV (%) / method No correction Di+MoCo
IVIM CV Slow diff. (D) [%] 16.98+/−4.63 14.72+/−5.52
CV Fast diff. (D*) [%] 58.15+/−28.19 46.91+/−9.79
CV Perf. Frac, (f) [%] 34.62+/−15.25 30.49+/−7.34
nRMSE 0.30+/−0.05 0.17+/−0.05
DTI CV MD [%] 6.50+/−1.25 3.28+/−1.17
CV FA [%] 14.51+/−1.67 12.09+/−2.31
nRMSE 0.35+/−0.06 0.12+/−0.02