. 2021 May;11(5):1817–1827. doi: 10.21037/qims-20-740

Table 3. Objective image quality.

Variables Study group 1, standard CT Study group 2, split-filter DECT P value
Pulmonary trunk (±SD)
   CT attenuation (HU) 428.2 (±26.5) 344.5 (±12.9) 0.013
   SNR 16.1 (±4.6) 27.0 (±8.2) <0.001
   CNR 13.9 (±4.6) 22.5 (±8.0) <0.001
Right lower lobe (±SD)
   CT attenuation (HU) 366.4 (±29.0) 327.3 (±13.3) 0.216
   SNR 14.4 (±6.2) 25.6 (±11.9) <0.001
   CNR 12.2 (±5.9) 21.2 (±11.4) <0.001
Left upper lobe (±SD)
   CT attenuation (HU) 365.3 (±36.9) 305.1 (±14.5) 0.031
   SNR 12.3 (±7.6) 23.7 (±12.9) <0.001
   CNR 10.4 (±7.0) 19.2 (±11.8) <0.001
Descending aorta (±SD)
   CT attenuation (HU) 223.9 (±22.9) 273.9 (±11.5) 0.001
   SNR 9.9 (±4.5) 24.2 (±6.8) <0.001
   CNR 7.4 (±4.4) 19.3 (±6.4) <0.001

Values provided as averages and standard deviations. DECT, dual-energy computed tomography; SNR, signal-to-noise ratio; CNR, contrast-to-noise ratio; SD, standard deviation.