Table 24. Multi-analyser comparability: possible materials to be used.
Recommended practice | Reanalysis of fresh patient samples on the different systems at a frequency defined by the laboratory and statistical validation according to standard ISO 5725-6:1994 (24). |
---|---|
Acceptable practice | IQC within acceptable limits for all systems EQA tested on all analysers with laboratory interpretation by the laboratory (statistical evaluation based on ISO 5725-6:1994) or by the EQA providers (provided that the frequency of these EQAs is sufficient) |
Unacceptable practice | Different acceptable limits on analysers. Compare only by looking at the average of normal patients (unable to quickly detect a discrepancy in one of the systems). |
IQC – internal quality control. EQA – external quality assessment. |