Table 1:

Image noise measured from 2 mm and 5 mm image before PKAID denoising, and from images after PKAID denoising using a prior image with slice thicknesses of 5, 8 and 10 mm, and relative change of noise (“+” indicates increase of noise) compared to the clinically used 5 mm image.

5 mm Original 2 mm 2 mm + PKAID (5 mm prior) 2 mm + PKAID (8 mm prior) 2 mm + PKAID (10 mm prior)
Noise(HU) 4.40 5.85 4.50 3.95 3.74
Relative Change - +36.6% +2.5% −11.4% −16.7%