Skip to main content
. 2021 Apr 16;293:114165. doi: 10.1016/j.jviromet.2021.114165

Table 1.

Clinical performance of the compared four PCR kits.

Molecular assay Final result1
PPA NPA
Positive Negative
Charité Positive 31 0 47.69 % 100 %
Negative 34 54
1copy Positive 59 5 90.77 % 90.74 %
Negative 6 49
Genesig Positive 64 1 98.46 % 98.15 %
Negative 1 53
DNA Technology Positive 60 1 92.31 % 98.15 %
Negative 5 53
1

The final result of a sample was positive when at least two of the four kits yielded a positive result.