Table 1:

Comparison of the temperature changes measured using FO sensor and MRT, and the corresponding SAR values calculated from these. The SAR values calculated from the MRT measurements were averaged over ROIs placed 2 mm from the tip of the FO sensors, rather than averaged over 10 g of phantom material. The SAR10g predicted by the 7T system and displayed on the console was 19.9 W/kg.

FO Sensor / ROI number FiberOptic sensor measurement MR Thermometry measurement Difference in dT/dt (FO sensor vs. MRT)
dT/dt [°C/min] SAR [W/kg] dT/dt [°C/min] SAR [W/kg] °C/min [95% CI]
1 0.237 14.22 0.252 15.12 −0.015 [−0.045, 0.016]
2 0.118 7.08 0.164 9.84 −0.046 [−0.056, −0.036]
3 0.020 1.20 0.039 2.34 −0.019 [−0.035, −0.003]
4 0.229 13.74 0.227 13.62 0.002 [−0.007, 0.011]