Table 3 ∣.
Troubleshooting table
| Step | Problem | Possible reason | Solution |
|---|---|---|---|
| 5 | All the grids have floated. | Glow discharge is not sufficient to cause enough hydrophilicity, or grids are not used within a suitable time (30 min) after the glow discharge. | Glow discharging of grids should be done within 30 min before usage. Verify grow discharge condition. |
| 12 | The cells do not adhere properly. | Glow discharging is not sufficient. The cells need another type of extracellular matrix or higher density |
See above. Optimize cell culture conditions. |
| 31 | Cryo-stage/Cryo-shield temperatures rise significantly above - 170 °C | N2 gas flow for cryo-shield and cryo-stage decreases significantly lower than needed. | Adjust the N2 gas flow at the source. If the same source of N2 gas is used for both the preparation station and cooling the cryo-stage, the prolonged use of N2 gas at the preparation station (e.g., for drying the preparation station) may sequester some N2 gas and reduce its flow rate into the cryo-stage and cryo-shield. A leak of cold N2 gas inside/along the heat exchanger. Check all the tubing connections. Unusual frost observed in the heat exchanger often indicates a location of cold N2 gas leak. |
| 57 & 59 | Ice contamination on top of the sample before milling | Contaminated liquid N2 during plunge freezing, clipping, and sample loading If the samples are observed by cryo-fluorescence microscopy, some degree of contamination should be accepted. Vacuum issues during the transfer |
Reduce room humidity Dry liquid N2 dewars before filling with fresh liquid N2 Filter liquid N2 with coffee filters Wear a mask to prevent contamination from the breath Carefully dry tools before re-use Check and clean valves and O-rings |
| 57 & 59 | Many broken EM grid meshes (Fig. 9a) | Damage during seeding cells on grids, plunge freezing or clipping | Evaporate a thin carbon layer on top of Quantifoil grids using carbon evaporator to make the substrate stronger Consider Quantifoil grids with SiO2 film known to provide more robust and stiffer support89. Manipulate grids more gently. Careful and gentle clipping, e.g., approach the EM grid in the FIB-AutoGrid with the C-ring Insertion Tool vertically; do not press the C-ring Insertion Tool vigorously onto the FIB-AutoGrid while pushing the C-ring down. |
| 57 & 59 | No cells on the EM grid | Insufficient glow discharging of grids Cell concentration too low |
Use glow discharged grids within 30 min Optimize parameters for glow discharge system Optimize cell concentration during seeding on the grids |
| 64 | When SEM and FIB do not show the spot on the grids even at the eucentric height. | The microscope is not well-aligned. | Reset beam shifts both in SEM and FIB. Use beam shift to adjust digitally. If still not possible, it is highly likely the FIB column position/angle needs to be adjusted mechanically. |
| 70 | Identification of cells on grids hindered after Pt Deposition | Thick Pt deposition. Check the thickness by the shape of the shadow of the protruding features. Pt deposition occurs at a slight angle, thus leaving behind a shadow for protruding features. |
Lower the Pt deposition time Changing the grid-needle distance. Changing the temperature of the GIS is also, not recommended. |
| 66-73 | FIB image moves | GIS Pt layer is not yet cured. Too thick ice on the entire sample Ice contamination along the focused ion beam path. The grid bar is along the FIB beam path. If temporal, it could be due to the change of stage temperature |
Use FIB to bake (Fig. 8b) Prepare a new sample. Increase the milling angle, if possible. Select a different area for milling. Increase milling angle Check stage temperature. |
| 66-73 | Milling takes an extremely long time | Platinum layer is too thick Milling the grid bar |
Reduce the thickness of the platinum layer by reducing the application time and/or increasing the GIS sample-to-needle distance Adjust milling position. Get familiar with how grid bars look and stop milling |
| 66-73 | Lamella bending/moving while milling | Charging is likely causing the beam-induced movement of the lamella Shrinkage of the EM grid support film at cryogenic temperatures |
Apply platinum layer Increase platinum layer thickness Carefully supervise milling process and compensate sample movement with ion beam shifts. Make micro-expansion joints84 |
| 66-73 | Lamella breaks while milling | Lamella too thin Beam-induced movement Shrinkage of the EM grid support film at cryogenic temperatures |
Make a thicker lamella See “Lamella moves while milling” Make micro-expansion joints84 |
| 99 | Vacuum pressure failure during the microscope warm-up | As water molecules frozen onto cold surfaces (e.g., cryo-shield) releases upon warming up, the pressure rises | Immediately activate pumping of the Microscope chamber. |
| 104-105 | Ice contamination on top of the milled lamellae | Contaminated grids Contaminated liquid N2 during transfers Vacuum issues during transfer |
Reduce contamination during grid preparation See “Ice contamination on top of the sample before milling” Check and clean valves and O-rings |
| 104-105 | Curtains | No or too thin platinum layer Crystalline ice on the surface before Pt coating The FIB current too high during milling The materials of significantly different density are present in the biological materials |
Increase the thickness of the platinum layer by increasing the application time and/or reducing the GIS sample-to-needle distance Minimize ice contamination Use lower FIB current to make smoother lamellae Make sure to complete material milling at each step before proceeding to the next milling. |
| 104-105 | The back part of the lamella is significantly thicker than front part (close to Pt layer). | Make sure to complete material milling at each step before proceeding to the next milling. Use different angles (plus/minus 0.5 °) for final milling step83 |
|
| 104-105 | Crystalline ice in the samples | Sample too thick | Increase blotting time Decrease cell concentration on the grid. If cells are too large for plunge freezing, this protocol might not suitable |
| 104-105 | “Leopard ice” on top of the lamellae | Shuttle/Stage temperature issue | Monitor stage top and shuttle temperature with thermocouples. Check the N2 flow of the heat exchanger. Inspect and clean surfaces of the upper and lower part of the cryo-stage and cryo-bearing. Retrieve the prepared EM grid from the microscope chamber as soon as possible after milling |
| 106 | Sample not aligned with milling direction 90° in regard to the tilt axis | Grids are not well-aligned in the shuttle for dual beam microscopy and/or in Titan cassette | Make sure to mark AutoGrid properly and orient them properly in dual beam and TEM. |
| 106 | Beam-induced motion while acquiring tilt series | Charging The lamella is too thin The lamella has cracks |
Consider coating lamella with Pt sputter coating Create thicker samples Gentle grid manipulation. See “Lamella bending/moving during milling” “Lamella breaks during milling” |
| 107 | Difficulties aligning tilt series | Lamellae are too thick to render good contrast The appearance of crystalline ice inside of samples Unstable Record/Tracking areas |
Make thinner samples. See below “Invisible amorphous ice layer on the top and/or the bottom of the reconstructed tomogram” Decrease pixel size Need to increase total electron dose See “Crystalline ice in the samples” Choose suitable area for tilt series acquisition. Gentle grid manipulation. See “Lamella bending/moving while milling” “Lamella breaks while milling” |
| 106-107 | Invisible amorphous ice layer on the top and/or the bottom of the reconstructed tomogram. The existing of this can be only confirmed by the observation that contaminated crystalline ice particles are not directly located on the biological layer. |
Redeposition during milling Contamination in Autoloader in Titan Krios |
Perform final fine milling just before retrieval of samples as mentioned in step 71. Perform “Cryo-cycle” |