Table 3.

User friendliness assessment of four COVID-19 antigen rapid diagnostic tests, adapted from SKUP/2018/114 protocol.

Mean of N = 3 questioned operators
Operation facilities BD Veritor™ SARS-CoV-2 Coris COVID-19 Ag Respi-strip Panbio™ COVID-19 Ag rapid test device SD Biosensor™ SARS-CoV-2 Rapid Antigen Test
To prepare the test Intermediate (1S 2I)a Intermediateb Intermediate (1S 2I)b Satisfactory (2S 1I)a
To prepare the sample Satisfactory Satisfactory Satisfactory Satisfactory
Application of specimen Satisfactory Satisfactory Satisfactory Satisfactory
Number of procedure step Satisfactory Satisfactory Satisfactory Satisfactory
Test design Satisfactory Unsatisfactory (2U, 1I)c Satisfactory Satisfactory
Reading of the result Satisfactory Difficultd Satisfactory Satisfactory
Sources of errors Satisfactory Intermediated Satisfactory Satisfactory
Hygiene when using the test Satisfactory Unsatisfactorye Satisfactory Satisfactory
Size and weight of the package Satisfactory (2S 1I)f Satisfactory Satisfactory Satisfactory
Storage conditions for tests, unopened package* 15–30°C 15–30°C 15–30°C 15–30°C
Storage conditions for tests, opened package* 15–30°C 15–30°C 15–30°C 15–30°C
Environmental aspects: waste handling* Special precautions Special precautions Special precautions Special precautions
Intended users* Health care personnel Health care personnel Health care personnel Health care personnel
Information in instruction in the insert
Preparations/Pre-analytic procedure Satisfactory Satisfactory Satisfactory Satisfactory
Specimen collection Satisfactory Intermediateg Satisfactory Satisfactory
Measurement procedure Satisfactory Satisfactory Satisfactory Satisfactory
Reading of result Satisfactory Satisfactory Satisfactory Satisfactory
Description of the sources of error Satisfactory Satisfactory Satisfactory Satisfactory
Help for troubleshooting Satisfactory Satisfactory Satisfactory Satisfactory
Readability/clarity of presentation Satisfactory Intermediateh Satisfactory Satisfactory (1I 2S)j
General impression Satisfactory Intermediatei Satisfactory Satisfactory (1I 2S)j
Measurement principle* Satisfactory Satisfactory Satisfactory Satisfactory
Available insert in ENG + FR + NL* Partlyk Partlyk Partlyl Partlyk
Time factors*
Required training time <2 h <2 h <2 h <2 h
Duration of preparations/Pre-analytical time <6 min <6 min <6 min <6 min
Duration of analysis 10–20 min >20 min 10–20 min 10–20 min
Stability of test, unopened package >5 months >5 months >5 months >5 months
Stability of test, opened package >30 days or disposable >30 days or disposable >30 days or disposable >30 days or disposable
Stability of quality control material unopened >5 months No QC provided >5 months No QC provided
Analytical quality control*
Reading of the internal quality control Satisfactory Unsatisfactorym Satisfactory Unsatisfactorym
Usefulness of the internal quality control Satisfactory Unsatisfactorym Satisfactory Unsatisfactorym
*

Objective informational items were filled by the principal investigator.

a

Caps of the buffer tubes difficult to manipulate.

b

Requires a tube rack.

c

The use of a strip in a closed tube with a very difficult capping was not considered practical for the operators.

d

Difficult reading through a closed tube although transparent.

e

Operators were forced to open the tubes to extract the strip in case of a doubt with the reading causing biosafety concern.

f

Oversized packaging compared to the number of test.

g

Lack of precise instruction.

h

Lack of clarity.

i

A quick reference guide would have been appreciated.

j

Small typo and dense content.

k

Only available in English.

l

Not available in Dutch.

m

Not provided.