Table 4.

Analytical performances of the Lumipulse® G SARS-CoV-2 Ag on target populations in the detection of SARS-CoV-2 using a categorization of the risk system.

Lumipulse® automated antigen detection
Certain (>20.27 pg/mL) Possible (1.32–20.27pg/mL) Unlikely (≤1.32pg/mL)
PCR result N Positive Lowest CT N Negative Positive Lowest CT N Negative Positive Lowest CT
Overall 279 14 3.89 46 39 7 15.7 219 214 5 22.61
Scheduled hospitalizations 93 1 12.73 6 4 2 15.7 86 85 1 26.04
Contacts 67 4 6.98 13 12 1 31.23 50 49 1 22.9
Health workers 119 9 3.89 27 23 4 19.92 83 80 3 22.61
 - With symptoms 67 8 3.89 13 13 - - 46 43 3 22.61