Abstract
The development of methods for the efficient and reliable separation and routine analysis of rare-earth elements (REEs), including samarium (Sm), proceeds to draw in the interest of the many researchers, attributable to the similar physical and chemical properties of these elements. Note that although the voltammetric determination of Sm has been described in the literature, thus far, no chemometric and voltammetric methods for the quantification of the element in its mixtures with other lanthanides in an acetonitrile solution have been reported. This work was aimed toward the advancement of a method for the detection of Sm in acetonitrile, the intended function of which was to obtain a selective current response of Sm by Differential Pulse Voltammetry, utilizing the Box-Behnken experimental design, to identify the best conditions for the determination. In particular, the three selected factors for the experiment, namely the potential range, amplitude modulation, and the deposition time, were found to have optimal conditions of −1.5 to +1.0 V, 0.075 V, and 60 s, consecutively. The optimal conditions were observed to result in a selective current response for samarium with a detection limit of 2.25 mg/L and a limit of quantitation of 7.50 mg/L. Furthermore, the accuracy was 98.70% and the precision was found to be 1.91% in relative standard deviation (RSD), while the recovery was found to be 98.70%.
Keywords: Samarium, Differential pulse voltammetry, Box–Behnken, Acetonitrile
Samarium, Differential pulse voltammetry, Box–Behnken, Acetonitrile.
1. Introduction
A group of chemical components that features the lanthanides group, yttrium, and scandium is called Rare Earth Elements (REE) [1]. Lately, the research about the REE has gained enormous attention due to the high application interest of the REE. These elements can be applied in various fields such as metallurgy, medical, nuclear technology, ceramic industry, and electronics [2, 3]. Samarium (Sm) is one of the 17 REE, which is ordinarily discovered in monazite and in bastnaesite. This element is not found in nature in a free state but in the form of complex compounds, and hence, a chemical analysis method is needed to separate this REE from its complex compounds [4, 5, 6, 7, 8]. Sm is relatively stable at high temperatures and has a thermoelectric efficiency of up to 1100 °C. This element is widely used in industrial fields, such as the computer, telecommunications, nuclear, and magnet industries [9, 10, 11, 12]. Besides, it can be used in the production of special catalysts, microwave and infrared equipment, lasers, and the nuclear energy industry [13, 14].
The sophisticated methods that are often used to determine the presence of Sm include HPLC offers a fast process, automatic and accurate analysis method, but in the process, this method is complicated and requires expensive equipment [15]. ICP–MS which has a high sensitivity for determining REEs, is simple, and only requires a small number of samples. However, spectral interference is the problem that must be solved in an ICP–MS analysis, because the molecular masses of REEs are adjacent [16]. Other methods such as ICP–OES, and XRF were used for analyzing many elements with sufficient sensitivity, but the performance of the tool is weak for REEs because of the complexity of the emission spectrum as well as the amount of interference from the major elements; therefore, the separation of the matrix by cation exchange is usually required [17]. These techniques must deal with complex dan costly maintenance and operation although have advantages in high selectivity and sensitivity. Therefore, an alternative method of analysis that is fast, efficient, sensitive, relatively inexpensive, and simple, and has a low detection limit is needed.
Voltammetry is an alternative analytical method that has the criteria for obtaining information about analytes by measuring currents in electrochemical cells as a function of potential. The current measured in a voltammetry analysis results from a redox reaction on the electrode surface [18, 19, 20]. The data obtained can be used for qualitative and quantitative analysis. Qualitative information is obtained from the current peak that appears on the voltammogram, which is the fingerprint of a particular analyte, while quantitative information is obtained from the area ratio of the voltammogram, which shows the composition ratio of the identified analyte [21].
The use of organic solvents as the supporting electrolyte solutions for metal electrodeposition continues to attract interest because it provides an alternative pathway to the electrochemical process. Organic solvents can accelerate the solubilization of and the reaction for metal ions [22, 23, 24]. Organic solvents that can be used include dimethylformamide (DMF), dimethyl sulfoxide (DMSO), acetonitrile (AcN), tetrahydrofuran (THF), and dimethylacetamide (DMAc) [25]. Compare to the other organic solvent, acetonitrile is the most widely used organic solvent for electrochemical studies. It is considered to be an excellent medium for the electrooxidation process due to has a high permittivity value, ε = 36, appropriate for a metal deposition because of a wide range of potential value, -3.45 V–2.35 V, [26, 27] and a relatively high dielectric constant; therefore, it dissolves electrolyte salts and can be mixed with ionic liquids to achieve very high ionic conductivity in the electrodeposition of metal ions [28, 29, 30].
Most of the analytical methods face the problem of time and high reagent consumption. An approach by the chemometric method can be used to solve those problems. This method assessed the significant factor that will influence the response. Amongst the tools for the analysis of chemometric, the Box-Behnken is considered effective to assign the correlation between the results of the response and the relevant factors by employing a sequence of experiments to provide the best responses [31, 32, 33, 34, 35, 36]. The experimental design technique mutual combination with an electroanalytical method was very promising for the future development of the determination of Sm and other REEs in water [9, 20, 37, 38].
Based on this background, this research was intended to develop a differential pulse voltammetry method using acetonitrile as a solvent for determining Sm levels. The Box-Behnken design technique was utilized to evaluate the optimal conditions of the experiments. The acetonitrile solvent effect for determining the presence of Sm(III) and for electrodeposition of the metal ions was investigated as well.
2. Materials and methods
2.1. Materials
The materials utilized in this research were distilled water, acetonitrile (Merck), samarium oxide (Sm2O3, 99.9%, Sigma Aldrich), dysprosium oxide (Dy2O3, 99.9%, Sigma Aldrich), europium oxide (Eu2O3, 99.9%, Sigma Aldrich), gadolinium oxide (Gd2O3 99.9%, Sigma Aldrich), and nitric acid 65% (Merck).
2.2. Apparatus
The tools and instruments utilized in this research included Ag/AgCl electrodes (eDAQ), Pt working electrodes (Antam), Pt wire for counter electrode (Antam), potentiostat (Metrohm® μAutolab), the Minitab 17.1 program, the ANOVA 7.0.0 program, analytical balance (Sartorius), Scanning Electron Microscope (SEM) JEOL JSM-7500F, and Hitachi TM3030 SwiftED3000 for EDX measurement.
2.3. Preparation of stock solution Sm2O3 1000 mg/L
The 1000 mg/L Sm2O3 was prepared by dissolving 0.2898 g Sm2O3 in a portion of 65% HNO3 followed by stirring and heating until homogenous. Into the solution, a portion of 250 mL distilled water was added and followed by stirring. A Sm2O3 stock solution was diluted using acetonitrile to prepare various concentrations of Sm.
2.4. Background current measurement
The current response of acetonitrile was measured by differential pulse voltammetry under the specific conditions: a potential of deposition −1.5 V, time of deposition 60 s, range of potential −1.5 V to +1.0 V, amplitude modulation of 0.05 V, and scanning rate of 0.05 V/s.
2.5. Sm current measurement
The current response of a 10.0 mg/L Sm solution was examined by differential pulse voltammetry under the specific conditions: a potential of deposition −1.5 V, time of deposition 60 s, range of potential −1.5 V to +1.0 V, amplitude modulation of 0.05 V, and scanning rate of 0.05 V/s. The measurement was repeated for: 15.0, 20.0, 25.0, and 30.0 mg/L.
2.6. Electrodeposition of Sm in acetonitrile by cyclic voltammetry
The deposition of 30.0 mg/L Sm solutions, in 25% and 100% acetonitrile, onto the Pt electrode was conducted by cyclic voltammetry under the specific conditions: a potential of deposition −1.5 V, time of deposition 60 s, range of potential −1.5 V to +1.0 V, amplitude modulation of 0.05 V, and scanning rate of 0.05 V/s.
2.7. Sm current response in 25% and 100% acetonitrile by differential pulse voltammetry
The current response of 30.0 mg/L Sm solution in 25% and 100% acetonitrile were examined was conducted using differential pulse voltammetry under the following conditions: a potential of deposition −1.5 V, time of deposition 60 s, range of potential −1.5 V to +1.0 V, amplitude modulation of 0.05 V, and scanning rate of 0.05 V/s.
2.8. Surface morphology analysis of platinum by SEM and EDX
The surface morphology of Pt (around 1 cm), before and after deposition with Sm, was characterized using SEM (JEOL JSM-7500F) and EDX (Hitachi TM3030 SwiftED3000).
2.9. Box–Behnken experimental design
The current responses of 30.0 mg/L Sm solution for the selected factors: deposition potential, deposition time, and amplitude modulation by utilizing differential pulse voltammetry were recorded. The Box–Behnken experimental was used to optimized the selected factors. The design was set to three levels (−1, 0, +1) with up to 15 measurements conducted using the Minitab program 17.1 [38].
2.10. Sm, Eu, Gd, and Dy current responses under the optimum conditions
The current responses of a 40.0 mg/L Sm, 0.1 mg/L Eu, 5.0 mg/L Gd, and 3.0 mg/L Dy in acetonitrile were observed by utilizing differential pulse voltammetry under the optimum conditions: a potential of deposition −1.5 V, time of deposition 60 s, range of potential −1.5 V to +1.0 V, amplitude modulation of 0.075 V, and scanning rate of 0.05 V/s.
2.11. Calibration curve of Sm
The current response of 10.0, 15.0, 20.0, 25.0, and 30.0 mg/L Sm solution were measured by differential pulse voltammetry under the optimum conditions a potential of deposition −1.5 V, time of deposition 60 s, range of potential −1.5 V to +1.0 V, amplitude modulation of 0.075 V, and scanning rate of 0.05 V/s.
3. Results and discussion
3.1. The acetonitrile and samarium current response
The differential pulse voltammetry was used to investigate the current response of acetonitrile and Sm in acetonitrile. These experiments were aimed to investigate whether the acetonitrile solvent shows a characteristic peak under the specific conditions. As shown in Figure 1, the acetonitrile solvent did not produce any current response, while the characteristic peak near −0.35 V is shown as the response of 10 mg/L Sm in acetonitrile. The reduction reaction of Sm could be expressed as follows:
Figure 1.
Current Response of acetonitrile and 10 mg/L Sm in acetonitrile.
The experiment was then conducted for investigated the current response of Sm in 25%, and 100% of acetonitrile solvent.
3.2. The current response of Sm in 25% and 100% acetonitrile and surface characterization of Pt electrodes
The electrochemical response investigation of Sm in 25% and in 100% acetonitrile was conducted by utilizing the differential pulse voltammetry and the cyclic voltammetry. The effect of the acetonitrile concentration on the deposition of Sm on the Pt electrode surface was evaluated in this experiment. The voltammograms of this experiment is presented in Figure 2.
Figure 2.
Differential pulse voltammogram (1) and cyclic voltammogram (2) profile of 30 mg/L Sm in (a) 100% acetonitrile and (b) 25% acetonitrile.
As shown in Figure 2(1), the differential pulse voltammogram of 30 mg/L Sm showed that in 100% acetonitrile, the current response was higher than that in 25% acetonitrile, which was probably due to the more facile adsorption of the metal on the platinum surface in a pure organic solvent; therefore, the Sm could be easily deposited on the electrode surface. In contrast, the higher water concentrations make it more difficult to deposit Sm on Platinum Electrode Surface. The investigation using cyclic voltammetry during the deposition of Sm, as shown in Figure 2(2), showed that in the case of 100% acetonitrile, the potential standard reduction E0 of Sm shifted toward a more cathodic potential. This indicates that the oxidation of Sm takes place more easily in 100% acetonitrile than in 25% acetonitrile.
The Pt surface was imaged before and after the deposition of Sm on the electrode surface by utilizing a Scanning Electron Microscope. These images are shown in Figure 3. As shown in Figure 3A, the image of bare Pt surface, whilst the morphology of the Pt surface after deposition of Sm in 25% and 100% acetonitrile are shown in Figures 3B and C. As shown in Figure 3B, there are a few 1 μm white aggregate particles on the Pt surface after deposition of Sm in 25% acetonitrile. Moreover, the white aggregate particles were found to be highly distributed on the platinum electrode surface after the deposition of Sm in 100% acetonitrile, as shown in Figure 3C. This result indicated that the concentration of acetonitrile solvent considerably influenced the deposition process of Sm ions on the platinum electrode.
Figure 3.
Pt electrode surface images (4300× magnification) before deposition(a), after deposition of Sm in 25% acetonitrile (b), and 100% acetonitrile (c).
Furthermore, the presence of Sm deposited on the Pt surface was supported by the EDX measurement result, as shown in Figure 4. The EDX spectrum presented in Figure 4A indicated the existence of Sm on the platinum surface with the %atomic of 0.013 after the deposition in 25% acetonitrile. The result of EDX spectra of the platinum surface after deposition of Sm in 100% acetonitrile shows the %atomic of 0.086, as shown in Figure 4B. This result confirmed that increasing the concentration of acetonitrile caused more Sm ions to deposit onto the platinum electrode.
Figure 4.
EDX spectrum of the Platinum Electrode Surface after deposition of (A) Sm in 25% acetonitrile and (B) Sm in 100% acetonitrile.
3.3. The determination of the optimum conditions
The optimum conditions for the measurement of Sm in acetonitrile were determined by the Box–Behnken method. The potential of deposition, amplitude modulation, and time of deposition was considered as the main factors that significantly influenced the determination of the optimum conditions of differential pulse voltammetry for the Sm analysis. The scanning potential was conducted from the negative potential (initial) to the positive potential (final). The given potential increased linearly, and the current was measured twice: before the pulse rose (given) and after the pulse fell (falls). Therefore, the deposition potential was chosen as a factor that might influence the Sm analysis. Furthermore, amplitude modulation affected the resolution (separation) and the sensitivity of the measurement. A large-amplitude modulation value would produce a large response. However, it would expand the peak and decrease the potential resolution. The optimum deposition time was the time at which the maximum peak current reduction signal was observed. Therefore, amplitude modulation and deposition time were also chosen as factors that might affect the Sm analysis.
The Box-Behnken analysis of the optimum condition for Sm in acetonitrile by differential pulse voltammetry was performed after determining the factors that potentially influenced the Sm analysis. The three levels were chosen which are high level (+), low level (-), and middle level (0), where the values are shown for each level, as presented in Table 1. These levels were generated from the Sm analysis without optimization, which could affect the voltammetry measurement.
| (1) |
Table 1.
The level of analysis and selected factors of Sm by differential pulse voltammetry.
| Symbol | Factor | Level |
||
|---|---|---|---|---|
| −1 | 0 | +1 | ||
| X1 | Deposition Potential/V | −1.0 | −1.5 | −2.0 |
| X2 | Deposition Time/s | 120 | 90 | 60 |
| X3 | Amplitude Modulation/V | 0.075 | 0.05 | 0.025 |
The results of the current response were obtained and processed to obtain the coefficient of the response function in Eq. (1).
As shown in Eq. (1), the X1 (deposition potential), X2 (amplitude modulation), and X3 (deposition time) were considered as the factors that had a positive or significant influence on the Sm analysis by the differential pulse voltammetry method. If the response coefficient was positive (+), the presence of these factors would increase the current response. A negative coefficient (−) indicated that the presence of these factors would decrease the current response. As shown in Eq. (1), the coefficient value of X1 (deposition potential) was negative, which implied a negative response in all three levels (−1, 0, +1). In contrast, the coefficient value of X2 and X3, i.e. amplitude modulation and deposition time, respectively, were positive, indicating the optimum current response that could be used in the Sm analysis will be increased due to the amplitude modulation and the time of deposition by utilizing differential pulse voltammetry.
From the analysis of the current response obtained from the differential pulse voltammetry, we obtained the ANOVA results presented in Table 2, which explained the variability of the data. The p-value data of each factor was obtained from the ANOVA. A model that in accordance with the linear model is indicated by a p-value of less than 0.05, which implied that a single variable showed a linear effect. The interaction effect is showed from a combination of two variables, and to determine the significance of each variable, the p-value was used. The independent variable which had no significant effect is indicated by the p-value of more than 0.05.
Table 2.
ANOVA results.
| Source | F-Value | P-Value |
|---|---|---|
| Model | 2.79 | 0.135 |
| Linear | 6.18 | 0.039 |
| Deposition Potential (X1) | 0.46 | 0.526 |
| Amplitude Modulation (X2) | 11.97 | 0.018 |
| Deposition Time (X3) | 6.10 | 0.056 |
| Square | 0.43 | 0.738 |
| Deposition Potential ∗ Deposition Potential (X1 ∗ X1) | 0.03 | 0.878 |
| Amplitude Modulation ∗ Amplitude Modulation (X2 ∗ X2) | 0.15 | 0.711 |
| Deposition Time ∗ Deposition Time (X3 ∗ X3) | 1.06 | 0.349 |
| Two-Way Interaction | 1.76 | 0.271 |
| Deposition Potential ∗ Amplitude Modulation (X1 ∗ X2) | 0.10 | 0.761 |
| Deposition Potential ∗ Deposition Time (X1 ∗ X3) | 0.95 | 0.374 |
| Amplitude Modulation ∗ Deposition Time (X2 ∗ X3) | 4.22 | 0.095 |
| Lack of Fit | 0.24 | 0.865 |
The p-value of the deposition potential (X1) was more than 0.05, as shown in Table 2. This result indicated that although we obtained a negative response from Eq. (1), it would not significantly influence the measurement result. Therefore, we kept applying the deposition potential of −1.5 V. The p-value of the lack of fit test was obtained as 0.865, which was greater than 0.05; thus, it concluded that the resulting linear model was appropriate. This lack of fit signified a deviation or inaccuracy with the linear model, and tests were conducted to detect whether the linear model was appropriate.
Moreover, based on the Minitab 17.1 program, the number of the conducted experiment for the three levels and three parameters was 15. The corresponding results are presented in Table 3. As shown in Table 3, the condition of the deposition potential of −1.5 V, deposition time of 60 s, and amplitude modulation of 0.075 V showed the high current response of 30 mg/L Sm in acetonitrile. This result was consistent with the result from the Box–Behnken method. Therefore, it concluded that the optimum conditions of the measurement were as follows: the potential of deposition −1.5 V, time of deposition 60 s, and amplitude modulation of 0.075 V.
Table 3.
The current response of Sm 30 mg/L in acetonitrile.
| Run | Factor |
Response/μA | |||
|---|---|---|---|---|---|
| Deposition Potential | Deposition Time | Amplitude Modulation | |||
| 1 | −1.5 V | 120 s | 0.075 V | 1.0646E−06 | |
| 2 | −1.5 V | 90 s | 0.050 V | 4.4952E−07 | |
| 3 | −1.0 V | 90 s | 0.025 V | 7.1832E−07 | |
| 4 | −1.0 V | 60 s | 0.050 V | 1.4863E−06 | |
| 5 | −1.5 V | 90 s | 0.050 V | 1.4640E−06 | |
| 6 | −1.5 V | 90 s | 0.050 V | 1.1316E−06 | |
| 7 | −1.0 V | 90 s | 0.075 V | 1.4838E−06 | |
| 8 | −2.0 V | 60 s | 0.050 V | 1.7548E−06 | |
| 9 | −1.5 V | 120 s | 0.025 V | 6.2622E−07 | |
| 10 | −1.0 V | 120 s | 0.050 V | 1.1198E−06 | |
| 11 | −1.5 V | 60 s | 0.075 V | 2.4401E−06 | |
| 12 | −2.0 V | 120 s | 0.050 V | 6.4522E−07 | |
| 13 | −2.0 V | 90 s | 0.075 V | 1.0978E−06 | |
| 14 | −1.5 V | 60 s | 0.025 V | 4.3647E−07 | |
| 15 | −2.0 V | 90 s | 0.025 V | 5.7724E−07 | |
3.4. Comparison of the responses of Sm, Eu, Gd, and Dy under the optimum conditions
The differential pulse voltammetry method was utilized to investigate the electrochemical responses of the Sm, Eu, Gd, and Dy under the optimum conditions. In nature, the monazite content of each REE (Sm, Eu, Gd, and Dy) is different from that of the others. The Sm content in the monazite of the Bangka and Belitung Islands is 4.21%, while the Eu content is 0.01%, the Gd content is 0.58%, and the Dy content is 0.37% [41, 42]. On the basis of this fact, we performed the measurement of different concentration ratios of Sm, Gd, Dy, and Eu. The Sm concentration used in this study was 40.0 mg/L, the Gd concentration was 5.0 mg/L, the Dy concentration was 3.0 mg/L, and the Eu concentration was 0.1 mg/L. The experiment result is shown in Figure 5.
Figure 5.
The individual current response of Sm 40.0 mg/L, Gd 5.0 mg/L, Dy 3.0 mg/L, and Eu 0.1 mg/L, and the mixture of Sm, Gd, Dy, and Eu under the optimum conditions by differential pulse voltammetry.
As shown in Figure 5. The voltammogram shows a relatively no response to the lowest current response for individual Gd, Dy, and Eu in acetonitrile. In contrast, the presence of Sm shows a high current response. The voltammogram of the mixture of Sm, Eu, Dy, and Gd shows the characteristic peak is similar for Sm. This result indicated that in the mixture with Eu, Dy, and Gd, the presence of Sm in acetonitrile can be identified.
Moreover, the observation of the electrochemical response of the individual Sm, Gd, Dy, and Eu and also its mixture under the same concentration (30 mg/L) was conducted by utilizing the differential pulse voltammetry. The voltammogram result is presented in Figure 6.
Figure 6.
The individual current response of Sm 30.0 mg/L, Gd 30 mg/L, Dy 30 mg/L, Eu 30 mg/L, and the mixture of Sm, Gd, Dy, and Eu under the optimum conditions by differential pulse voltammetry.
The responses of the mixture of Sm, Gd, Dy, and Eu (30 mg/L) is relatively similar to that of individual Sm, as shown in Figure 6. Therefore, this result confirmed that the presence of Sm can be identified in the mixture with Gd, Eu, and Dy.
3.5. Calibration curves, detection limits, and quantification limits
The current peak of Sm under various concentrations: 10.0, 15.0, 20.0, 25.0, and 30.0 mg/L in a 100% acetonitrile, were recorded and used to prepare the calibration curves for Sm. These calibration curves were utilized for a comparison of the analyte concentrations versus the responses given in the form of linear regression. The optimum conditions from the Box-Behnken were applied for the Sm current response measurement.
From the result, as shown in Figure 7, the increasing concentration is proportional to the current peak due to of the large number of electroactive analyte ions (Sm3+), which were reduced or deposited on the Pt electrode at -0.35 V. This was consistent to the Rendless-Sevcik equation, in which the current is correlated with the analyte concentration. The given negative potential promoted the Pt electrode to have a more negative charge. The Sm3+ ions on the surface of the electrode were reduced (capturing electrons) to Sm2+. As a consequence, this decreased the concentration of the Sm ions on Pt electrode surface, and the Sm ions far from the electrode (the concentration was greater than on Pt electrode surface) moved toward the Pt electrode. We assumed that the electron movement mechanism due to the different concentrations caused the current diffusion, which was equal to the analyte concentration.
Figure 7.
(A) Differential Pulse Voltammogram and (B) Cyclic Voltammogram of acetonitrile and variation concentration of Sm in acetonitrile from 10 to 30 mg/L (potential of deposition -1.5 V, range of potential −1.5 V to +1.0 V, amplitude modulation 0.075 V, time of deposition 60 s, and scanning rate of 0.05 V/s).
As shown in Figure 8, the linear regression of calibration curve was calculated to be y = 0.0966x − 0.5635 with R2 equal to 0.9963. This value suggested that Sm had a strong variable relationship between the response and the concentration.
Figure 8.
Calibration curve of Sm in acetonitrile from 10 to 30 mg/L.
We performed the experiments under the optimal condition and calculated the values of the limit of quantification (LoQ) and the limit of detection (LoD) [39, 40]. As shown in Table 4, the LoD value obtained was 2.25 mg/L, and the LoQ was 7.50 mg/L. The experiment for the determination of accuracy, precision, and recovery was also investigated. The results are presented in Table 5. The accuracy (for 30 mg/L Sm) was 98.70% while the precision values was 1.91% in RSD. The recovery of 30 mg/L Sm was 98.70% (see Table 6).
Table 4.
Current response of Sm for determination of LoQ and LoD.
| Sm/mg/L (x) |
Current/μA (y) |
||||
|---|---|---|---|---|---|
| 10 | 0.4495 | 0.4025 | 0.0470 | 0.0022 | |
| 15 | 0.8380 | 0.8855 | −0.0475 | 0.0022 | |
| 20 | 1.3218 | 1.3685 | −0.0467 | 0.0021 | |
| 25 | 1.8960 | 1.8515 | 0.0445 | 0.0019 | |
| 30 |
2.3350 |
2.3345 |
0.0005 |
0.0000 |
|
| 100 | 6.8404 | 6.8425 | −0.0022 | 0.0086 | |
| Mean | 20 | 1.3680 | 0 |
Table 5.
Current response of Sm for determination of precision, accuracy, and recovery.
| Sm/mg/L (x) |
Current/μA (y) |
Smexperiment/mg/L | Sb | KV | Precision/% RSD | Accuracy/% | |||
|---|---|---|---|---|---|---|---|---|---|
| 10 | 0.4495 | 10.4865 | −0.3723 | 0.1386 | 0.2565 | 0.3581 | 3.2979 | 3.30 | 91.41 |
| 0.4884 | 10.8892 | 0.0304 | 0.0009 | ||||||
| 0.5185 | 11.2008 | 0.3420 | 0.1169 | ||||||
| Mean | 10.8589 | ||||||||
| 15 | 0.8380 | 14.5083 | 0.0949 | 0.0090 | 0.0963 | 0.2194 | 1.5221 | 1.53 | 96.08 |
| 0.8439 | 14.5694 | 0.1560 | 0.0243 | ||||||
| 08046 | 14.1625 | −0.2509 | 0.0629 | ||||||
| Mean | 14.4134 | ||||||||
| 20 | 1.3218 | 19.5166 | −0.2460 | 0.0605 | 0.2057 | 0.3207 | 1.6226 | 1.63 | 98.81 |
| 1.3343 | 19.6460 | −0.1166 | 0.0136 | ||||||
| 1.3806 | 20.1253 | 0.3627 | 0.1315 | ||||||
| Mean | 19.7626 | ||||||||
| 25 | 1.8960 | 25.4607 | 0.9907 | 0.9815 | 4.9645 | 1.5755 | 6.4386 | 6.44 | 97.87 |
| 1.8801 | 25.2961 | 0.8261 | 0.6824 | ||||||
| 1.6248 | 22.6532 | −1.8168 | 3.3007 | ||||||
| Mean | 24.4700 | ||||||||
| 30 | 2.2343 | 28.9625 | −0.6478 | 0.4197 | 0.6396 |
0.5655 |
1.9098 |
1.91 |
98.70 |
| 2.3213 | 29.8634 | 0.2530 | 0.0640 | ||||||
| 2.3350 | 30.0052 | 0.3948 | 0.1559 | ||||||
| Mean | 29.6014 | ||||||||
| Mean | 2.97 | 96.57 | |||||||
Table 6.
Comparison of the result with the previous work.
| Method | Result |
|---|---|
| A study of electroanalysis and experimental design combination method for detection of Sm-DTPA complex (Solvent: NH4Cl) [9] | LoD: 24.44 mg L−1 |
| LoQ: 91.53 mg L−1 | |
| Not selective for Sm, Eu and Dy | |
| An application of differential pulse voltammetry and experimental design method for determination of REE-DTPA complex (Solvent: NH4Cl) [20] | LoD: 27.11 mg L−1 |
| LoQ: 92.33 mg L−1 | |
| Not selective for Sm, Eu, and Dy | |
| A differential pulse voltammetry study for detection of Dy (III) in acetonitrile (Solvent: CH3CN) [38] | LoD: 0.64 mg L−1 |
| LoQ: 2.14 mg L−1 | |
| Not selective for Eu and Dy | |
| This work (Solvent: CH3CN) | LoD: 2.25 mg L−1 |
| LoQ: 7.50 mg L−1 | |
| Selective for Sm but not selective for Eu, Gd and Dy |
4. Conclusion
The results of this study indicated that Sm in acetonitrile could be deposited on platinum electrodes. Based on the Box–Behnken experimental design result, the optimum conditions for determining Sm by differential pulse voltammetry were as follows: potential of deposition -1.5 V, amplitude modulation 0.075 V, and time of deposition 60 s. The LoD obtained was 2.25 mg/L, and the LoQ was 7.50 mg/L. The accuracy was 96.70%, and the precision was 1.91% in RSD, while the recovery was 98.70%. This study suggested that the differential pulse voltammetry method could determine the presence of Sm in acetonitrile.
Declarations
Author contribution statement
Santhy Wyantuti: Conceived and designed the experiments; Performed the experiments; Analyzed and interpreted the data; Contributed reagents, materials, analysis tools or data; Wrote the paper.
Uji Pratomo: Performed the experiments; Analyzed and interpreted the data; Wrote the paper.
Lastri A. Manullang, Diana Hendrati: Performed the experiments; Analyzed and interpreted the data.
Yeni Wahyuni Hartati: Conceived and designed the experiments; Performed the experiments; Analyzed and interpreted the data; Wrote the paper.
Husein H. Bahti: Conceived and designed the experiments; Contributed reagents, materials, analysis tools or data; Wrote the paper.
Funding statement
Santhy Wyantuti was supported by Academic Leadership Grant Program, Padjadjaran University, and the Directorate of Research and Community Service through Superior Research of Higher Education (DRPM-PDUPT), Padjadjaran University (1827/UN6.D/LT/2020).
Data availability statement
Data associated with this study has been deposited at Penerapan Desain Box-Behnken pada Penentuan Kadar Sm (III) Tunggal dan Dalam Campuran Sm, Eu, Gd, Dy Menggunakan Pelarut Asetonitril Secara Voltammetri under the accession number http://repository.unpad.ac.id/frontdoor/index/index/docId/12088.
Declaration of interests statement
The authors declare no conflict of interest.
Additional information
No additional information is available for this paper.
References
- 1.Abre R.D., Morais C. Purification of rare earth elements from monazite sulphuric acid leach liquor and the production of high-purity ceric oxide. Miner. Eng. 2010;23:536–540. [Google Scholar]
- 2.de Vasconcellos M.E., Queiroz C.A.S., Abrao A. Sequential separation of the yttrium heavy rare earths by fractional hydroxide precipitation. J. Alloys Compd. 2004:405–407. [Google Scholar]
- 3.Davodi-Nasab P., Rahbar-Kelishami A., Safdari J., Hossein A. Evaluation of the emulsion liquid membrane performance on the removal of gadolinium from acidic solutions. J. Mol. Liquid. Iran. 2018;262:97–103. [Google Scholar]
- 4.Hurst C. IAGS; USA: 2010. China’s Rare Earth Elements Industry: what Can the West Learn. [Google Scholar]
- 5.Rodliyah I., Siti R., Tatang W. Extraction of rare earth metals from monazite mineral using acid method. Indonesian Mining J. 2014;18(1):39. [Google Scholar]
- 6.Taam I., Jesus C.S., Mantovano J.L., Gante V. Quantitative analysis of rare earths by x-ray fluorescence spectrometry. Int. Nuclear Atlantic Conf. 2013:24–29. [Google Scholar]
- 7.Schramm R. Use of x-ray fluorescence analysis for the determination of rare earth elements. Phys. Sci. Rev. 2016:1–17. [Google Scholar]
- 8.Smoli A., Stempin M., Howaniec N. Determination of rare earth elements in combustion ashes from selected Polish coal mines by wavelength dispersive x-ray fluorescence spectrometry. Spectrochim. Acta, Part B. 2016;116:63–74. [Google Scholar]
- 9.Wyantuti S., Pratomo U., Hartati Y.W., Hendrati D., Bahti H.H. A study of green electro-analysis conducted by experimental design method for detection of samarium as complex with diethylenetriaminepentaacetic acid (DTPA) AIP Conf. Proc. 2018 [Google Scholar]
- 10.Elsner H. Federal Institute of Geosciences and Natural Resource (BGR); Hannover: 2010. Heavy Minerals of Economic Importance: Assessment Manual. [Google Scholar]
- 11.Fontana D., Loris F. Separation of middle rare earths by solvent extraction using 2-ethylheksylphosphonic acid mono-2-ethylheksyl ester. J. Rare Earths. 2009;27(3):117–135. [Google Scholar]
- 12.Liu B.L., Lu M., Wang Z.I. A study of nanostructured gold modified glassy carbon electrode for the determination of trace Cr (IV). Indian Academy of Sciences. J. Chem. Sci. 2008;120:493–498. [Google Scholar]
- 13.Torkaman R., Moosavian M.A., Torab-Mostaedi M., Safdari J. Solvent extraction of samarium from aqueous nitrate solution by Cyanex 301 and D2EHPA. Hydrometallurgy. 2013;137:101–107. [Google Scholar]
- 14.Finlay I.G., Mason M.D., Shelley M. Radioisotopes for the palliation of metastatic bone cancer: a systematic review. Lancet Oncol. 2005;6(6):392–400. doi: 10.1016/S1470-2045(05)70206-0. [DOI] [PubMed] [Google Scholar]
- 15.Sakr M., Hanafi R., Fouad M., Al-Easa H., El-Moghazy S. Design and optimization of a luminescent Samarium complex of isoprenaline: a chemometric approach based on Factorial design and Box-Behnken response surface methodology. Spectrochim. Acta Mol. Biomol. Spectrosc. 2019;208:114–123. doi: 10.1016/j.saa.2018.09.061. [DOI] [PubMed] [Google Scholar]
- 16.Zawisza B., Pytlakowska K., Feist B., Polowniak M., Kita A., Sitko R. Determination of rare earth elements by spectroscopic techniques: a review. J. Anal. Atomic Spectrom. 2011;26(12):2373–2390. [Google Scholar]
- 17.Ardini F., Soggia F., Rugi F., Udisti R., Grotti M. Comparison of inductively coupled plasma spectrometry techniques for the direct determination of rare earth elements in digests from geological samples. Anal. Chim. Acta. 2010;678(1):18–25. doi: 10.1016/j.aca.2010.07.036. [DOI] [PubMed] [Google Scholar]
- 18.Ganjali M.R., Zare-Dorabei R., Norouzi P. Design and construction of a novel optical sensor for determination of trace amounts of dysprosium ion. Sensor. Actuator. B Chem. 2009;143(1):233–238. doi: 10.1016/j.jhazmat.2009.06.044. [DOI] [PubMed] [Google Scholar]
- 19.Skoog D.A., West D.M., Holler F.J. Cengage Learning; Boston: 2014. Fundamentals of Analytical Chemistry. [Google Scholar]
- 20.Wyantuti S., Pratomo U., Hartati Y.W., Hendrati D., Bahti H.H. Application of experimental design by differential pulse voltammetry for determination of rare earth elements as complexes with diethylenetriaminepentaacetic acid (DTPA) Int. J. Recent Technol. Eng. 2019;8:33–37. [Google Scholar]
- 21.Scholz F., Lange B. Abrasive stripping voltammetry — an electrochemical solid state spectroscopy of wide applicability. Trac. Trends Anal. Chem. 1992;11(10):359–367. [Google Scholar]
- 22.Neuróhr K., Pogány L., Tóth B.G., Révész Á., Bakonyi I., Péter I. Electrodeposition of Ni from various non-aqueous media: the case of alcoholic solutions. J. Electrochem. Soc. 2015;162(7):D256–D264. [Google Scholar]
- 23.Bourbos E., Karantonis A., Sygellou L., Paspaliaris I., Panias D. Study of Nd electrodeposition from the aprotic organic solvent dimethyl sulfoxide. Metals. 2018;8:803. [Google Scholar]
- 24.Mirzamohammadi S.H., Khorsand H., Aliofkhazraei M. Effect of different organic solvents on electrodeposition and wear behavior of Ni-alumina nanocomposite coatings. Surf. Coating. Technol. 2017;313:202–213. [Google Scholar]
- 25.Jung Y., Kim S., Kim B.-S., Han D.-H., Park S.-M., Kwak J. Effect of organic solvents and electrode materials on electrochemical reduction of sulfur. Int. J. Electrochem. Sci. 2008;3:566–577. [Google Scholar]
- 26.Kiss L. Electrooxidation of low-permittivity solvents in acetonitrile and solubility of trihexylamine in acetonitrile. J. Iran. Chem. Soc. 2020;17:67–71. [Google Scholar]
- 27.Elgrishi N., Rountree K.J., McCarthy B.D., Rountree E.S., Eisenhart T.T., Dempsey J.L. A practical beginner’s guide to cyclic voltammetry. J. Chem. Educ. 2018;95:197–206. [Google Scholar]
- 28.Koverga V.O. A new potential model for acetonitrile: insight into the local structure organization. J. Mol. Liq. 2017;233:251–261. [Google Scholar]
- 29.Suchacz B., Wesolowski M., Yu F., Wang X.X., Yang N., Compton R.G., Yu M., Dai S., Liu G., Shakil A.O., Lee W.M. Voltammetric quantitation of acetaminophen in tablets using solid graphite electrodes. Anal. Methods. 2016;8(16):3307–3315. [Google Scholar]
- 30.Yanez I. Hydrogen oxidation and hydrogen evolution on a platinum electrode in acetonitrile. ChemElectroChem. 2015:1612–1622. [Google Scholar]
- 31.Tarley C.R.T., Silveira G., dos Santos W.N.L., Matos G.D., da Silva E.G.P., Bezerra M.A., Miro M., Ferreira S.L.C. Chemometric tools in electroanalytical chemistry: methods for optimization based on factorial design and response surface methodology. Microchem. J. 2009;92:58–67. [Google Scholar]
- 32.Wiik H.E. Norwegian University of Science and Technology; Norway: 2014. Methods for Analyzing the 12 Run Plackett–Burman Design. [Google Scholar]
- 33.Montgomery D.C. John Willey & Sons; Singapore: 2001. Design and Analysis of Experiments. [Google Scholar]
- 34.Bezerra R. Response surface methodology (RSM) as a tool for optimization in analytical chemistry. Talanta. 2008:965–977. doi: 10.1016/j.talanta.2008.05.019. [DOI] [PubMed] [Google Scholar]
- 35.Elazazy M.S., El-Hamshary M., Sakr M., Al-Easa H.S. Plackett-Burman and Box-Behnken designs as chemometric tools for micro-determination of LOrnithine. Spectrochim. Acta Mol. Biomol. Spectrosc. 2018;193:397–406. doi: 10.1016/j.saa.2017.12.044. [DOI] [PubMed] [Google Scholar]
- 36.Morgan E. Published on behalf of ACOL by John Wiley & Sons; Chichester: 1997. Chemometrics: Experimental Design. [Google Scholar]
- 37.Wyantuti S., Pratomo U., Hartati Y.W., Anggraeni A., Bahti H.H. Fast and simultaneous detection of Sm, Eu, Gd, Tb and Dy using combination of voltammetry method and multivariate analysis. Res. J. Chem. Environ. 2018;22:302–306. [Google Scholar]
- 38.Wyantuti S., Pratomo U., Shauvina S., Hartati Y.W., Bahti H.H. Differential pulse voltammetry study for quantitative determination of dysprosium (III) in acetonitrile solution. Int. J. Renew. Energy Dev. 2021;10:191–199. [Google Scholar]
- 39.Miller J.N., Miller J.C., Miller R.D. seventh ed. Pearson Education Limited; Harlow: 2018. Statistics and Chemometrics for Analytical Chemistry. [Google Scholar]
- 40.Bard J.A., Faulkner L.R. second ed. John Wiley & Sons; 2001. Electrochemical Methods: Fundamentals and Applications. [Google Scholar]
- 41.Irawan S. Bandung; 2013. Rare Earth Element Mineral Processing in PT Timah Tbk: Workshop on Rare Earth Element Mineral Processing and Refining in Indonesia. [Google Scholar]
- 42.Suhkyat R. Seminar on Acceleration Attempts to Increase Domestic Minerals Added Value. 2013. Potential and distribution of Indonesia mineral resources and deposits. Indonesia. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data associated with this study has been deposited at Penerapan Desain Box-Behnken pada Penentuan Kadar Sm (III) Tunggal dan Dalam Campuran Sm, Eu, Gd, Dy Menggunakan Pelarut Asetonitril Secara Voltammetri under the accession number http://repository.unpad.ac.id/frontdoor/index/index/docId/12088.