Skip to main content
. 2021 Mar 24;25:e00218. doi: 10.1016/j.plabm.2021.e00218

Table 1.

Assay imprecision.

Method Patient Pools Mean Concentration (mg/dL) Total Imprecision (SD) Total Imprecision (%CV) Within Run (%CV) Between Run (%CV) Between Day (%CV)
Diazyme Low 12.4 0.31 2.5 2.1 1.2 0.8
High 50.2 0.98 2.0 1.4 0.0 1.4
Kamiya Low 8.2 0.43 5.2 4.7 0.0 2.3
High 28.6 0.86 3.0 2.9 0.0 0.8
MedTest Low 12.3 0.48 3.9 2.8 2.8 0.0
High 47.5 0.55 1.1 0.8 0.9 0.0
Randox Low 9.5 0.32 3.3 3.2 0.7 0.0
High 35.7 0.29 0.8 0.5 0.5 0.4
Roche Low 7.3 0.31 4.3 4.0 0.0 1.6
High 31.5 0.36 1.1 0.8 0.8 0.0