Table 2.
CNR improvement factor, KCNR, in small and large Catphan phantoms.
| Phantom size | GSS used? | 1D ASG | 2D ASG | ||
|---|---|---|---|---|---|
| R21 | R8P3 | R12P2 | R16P2 | ||
| Small | No | 1.05 ± 0.09 | 1.09 ± 0.11 | 1.31 ± 0.06 | 1.21 ± 0.15 |
| Yes | N/A | 1.10 ± 0.12 | 1.31 ± 0.06 | 1.21 ± 0.15 | |
| Large | No | 1.49 ± 0.42 | 1.87 ± 0.37 | 2.01 ± 0.38 | 1.95 ± 0.50 |
| Yes | N/A | 1.86 ± 0.37 | 2.01 ± 0.40 | 2.00 ± 0.50 | |