Skip to main content
. 2021 Apr 14;13(2):188–194. doi: 10.5114/jcb.2021.105287

Table 2.

Measurement uncertainty analysis

Variable Film (%) Farmer (%)
Scanning consistency 0.1 0.1
Film uniformity 1.0 0.0
ROI size 1.5 0.0
Calibration curve fit 3.5 0.0
Chamber calibration 0.0 3.0
Distance to source 4.0 6.0
Source strength 5.0 5.0
Exposure time 0.5 0.5
Repeatability 3.0 3.0
Total uncertainty 8.0 8.8