Table 2.

Dose distributions in HDR surface brachytherapy. Depths were calculated from the source position (not from the skin)

CTV size (cm2) CTV shape 5% dose depth (mm) 50% dose depth (mm) 90% dose depth (mm) 125% dose depth (mm)
10 × 10 PLANE CAX 138 27 10 5
Edge 130 21 9 6
Lateral 96 14 7 5
10 × 10 CAX30 CAX 136 32 10 5
Edge 110 16 8 5
Lateral 92 16 8 5
10 × 10 CAX60 CAX 139 38 11 4
Edge 104 15 6 4
Lateral 98 16 7 5
10 × 10 HCX CAX 139 47 12 4
Edge 99 14 6 4
Lateral 100 17 6 4
20 × 20 PLANE CAX 151 34 11 5
Edge 134 22 9 6
Lateral 101 15 7 5
20 × 20 CAX30 CAX 164 39 11 4
Edge 115 17 8 5
Lateral 105 15 7 5
20 × 20 CAX60 CAX 192 52 12 5
Edge 164 39 11 4
Lateral 180 47 11 4
20 × 20 HCX CAX 192 52 12 5
Edge 164 39 11 4
Lateral 180 47 11 4
d = 16 cm DS CAX 194 61 12 4
Edge 107 18 9 5
Lateral 98 16 7 4