Table VI.

Performances of Truenat and Xpert in comparison with culture among smear-negative samples

Truenat Culture Total Xpert Culture Total


Positive Negative Positive Negative
Positive 183 376 559 Positive 148 160 308
Negative 61 1226 1287 Negative 96 1442 1538
Total 244 1602 1846 Total 244 1602 1846
Sensitivity: 75% (69.1-80.3%) Sensitivity: 60.7% (54.2-66.8%)
Specificity: 76.5% (74.4-78.6%) Specificity: 90.0% (88.4-91.4%)

Sensitivity of Truenat and Xpert in detecting MTB among smear-negative samples in comparison with culture was 75 and 60.7% while specificities were 76.5 and 90.0%, respectively. The differences in sensitivity and specificity between Truenat and Xpert were significant (P<0.001)