Skip to main content
. 2021 Apr 22;16(4):e0249697. doi: 10.1371/journal.pone.0249697

Table 3. MPI-calculated velocities in stenotic phantoms.

Setting Flow rate [ml/min] Reference velocity [cm/s] FDHM velcoties using intrinsic correction factor [cm/s] Correction factor c [1] Absolute error [cm/s]/Absolute relative error FDHM velocities using correction factor from setting 2 Absolute error [cm/s]/Absolute relative error FDHM using correction factor from setting 3 Absolute error [cm/s]/Absolute relative error
1 151 ± 24.5 20 ± 2.2 19.8 ± 2.0 25.30 ± 0.19 0.2/1.2%
21.5 ± 1.8 25.84 ± 0.08 1.5/7.7%
19.0 ± 2.0 25.03 ± 0.14 1.0/5.1%
19.8 ± 2.0 25.32 ± 0.19 0.2/0.8%
2 680 ± 40.4 40 ± 3.6 42.7 ± 2.8 21.22 ± 0.13 2.7/6.8%
42.3 ± 2.9 21.18 ± 0.14 2.3/5.7%
34.9 ± 3.1 20.49 ± 0.09 5.1/12.9%
42.3 ± 2.9 21.18 ± 0.14 2.3/5.7%
42.4 ± 2.9 21.19 ± 0.14 2.4/6.0%
36.8 ± 3.1 20.69 ± 0.13 3.2/8.0%
3 205 ± 26.2 12 ± 1.7 12.9 ± 1.6 17.09 ± 0.32 0.9/7.1% 15.8 ± 1.1 3.8/31.6%
9.0 ± 1.5 16.00 ± 0.25 3.0/25.2% 11.8 ± 1.1 0.2/1.9%
12.9 ± 1.6 17.10 ± 0.32 0.9/7.5% 15.8 ± 1.1 3.8/32.0%
15.2 ± 1.4 17.51 ± 0.29 3.2/26.6% 18.2 ± 1.0 6.2/51.8%
16.1 ± 1.3 17.65 ± 0.27 4.1/34.4% 19.2 ± 0.9 7.2/59.9%
9.1 ± 1.5 16.04 ± 0.26 2.9/24.2% 11.9 ± 1.1 0.1/0.9%
4 410 ± 32.3 24 ± 2.5 22.7 ± 1.6 13.79 ± 0.06 1.3/5.6% 34.5 ± 1.1 10.5/43.7% 27.8 ± 1.3 3.8/15.7%
25.2 ± 1.6 14.28 ± 0.11 1.2/4.8% 37.0 ± 1.0 13.0/54.1% 29.8 ± 1.3 5.8/24.0%
24.3 ± 1.7 14.11 ± 0.17 0.3/1.1% 36.1 ± 1.1 12.1/50.4% 29.1 ± 1.3 5.1/21.1%

This table shows the velocities for the FDHM method in comparison to the reference values for the four prior specified settings (1: 4 mm diameter, no stenosis; 2: 6 mm diameter, no stenosis; 3: 6 mm diameter, stenosis; 4: 6 mm diameter, stenosis, doubled flow rate as in 3). At first, FDHM velocities were calculated using the cross-correlated correction factor of the respective measurement series, representing the iCF. In a next step, FDHM velocities for setting 3 and 4 were calculated by using the correction factor of setting 2 to evaluate feasibility of non-stenotic tubes being used as a calibration for stenotic ones. Finally, the correction factor of setting 3 was used to calculate the FDHM velocities in setting 4 in order to assess calibration viability using stenotic tubes. TD-method was not used due to inferior performance in comparison to FDHM.