Table 5.
Diagnostic performance characteristics of contrast-enhanced spectral mammography without delayed image and contrast-enhanced spectral mammography with delayed image
| CESM standard | CESM with delayed image | |
|---|---|---|
| Sensitivity (%)a | 97.8 | 93.7 | 
| Specificity (%)a | 80.0 | 92.4 | 
| PPV (%)a | 66.7 | 89.4 | 
| NPV (%)a | 98.9 | 95.5 | 
| Accuracy (%)a | 64.27 | 92.9 | 
| AUCb | 0.924 (0.883–0.964) | 0.969 (0.942–0.995) | 
| p-value | < 0.001 | |
– values in parentheses are raw numbers used to calculate percentages, b – values in parentheses are 95% CI, PPV – positive predictive value, NPV – negative predictive value, AUC – area under the curve