Skip to main content
. 2021 Apr 1;11(2):151–162. doi: 10.31661/jbpe.v0i0.2001-1061

Table 3.

Image Linearity (IL) measurements conducted on Vatech EzRay Premium Vex S100W(1) equipment.

P1 P2 D
1st measure of DM (mm) 7.99 ± 0.40 8.00 ± 0.40 7.99 ± 0.40
2nd measure of DM (mm) 8.00 ± 0.40 7.99 ± 0.40 7.99 ± 0.40
3rd measure of DM (mm) 8.00 ± 0.40 8.00 ± 0.40 8.00 ± 0.40
4th measure of DM (mm) 7.99 ± 0.40 7.99 ± 0.40 7.98 ± 0.40
Image Linearity (%) 0.06 ± 0.01 0.06 ± 0.01 0.13 ± 0.01