Table 4.
Quantitative quality control (QC) measurements conducted on all considered devices.
| Device | Rater | IU (GU) | HCSR (10-1 mm) | IL (%) |
|---|---|---|---|---|
| Kodak 210 | P1 | |17.79 ± 7.41| | 0.83 ± 0.05 | 0.07 ± 0.01 |
| P2 | |14.62 ± 6.27| | 0.83 ± 0.06 | 0.07 ± 0.01 | |
| D | |15.38 ± 6.25| | 0.90 ± 0.05 | 0.08 ± 0.01 | |
| Orix 70 | P1 | |18.07 ± 7.62| | 0.85 ± 0.10 | 0.05 ± 0.01 |
| P2 | |18.52 ± 7.83| | 0.70 ± 0.08 | 0.05 ± 0.01 | |
| D | |16.35 ± 6.84| | 0.86 ± 0.07 | 0.05 ± 0.01 | |
| Carestream CS 2100 | P1 | |5.58 ± 4.23| | 1.40 ± 0.06 | 0.06 ± 0.01 |
| P2 | |4.94 ± 3.54| | 1.43 ± 0.06 | 0.06 ± 0.01 | |
| D | |5.24 ± 3.76| | 1.30 ± 0.05 | 0.08 ± 0.01 | |
| Vatech EzRay Premium Vex S100W (1) | P1 | |6.79 ± 4.72| | 1.25 ± 0.05 | 0.06 ± 0.01 |
| P2 | |8.86 ± 4.86| | 1.25 ± 0.02 | 0.06 ± 0.01 | |
| D | |11.10 ± 4.90| | 1.23 ± 0.11 | 0.13 ± 0.01 | |
| Vatech EzRay Premium Vex S100W (2) | P1 | |7.42 ± 4.07| | 1.20 ± 0.05 | 0.07 ± 0.01 |
| P2 | |8.96 ± 4.64| | 1.30 ± 0.05 | 0.09 ± 0.01 | |
| D | |8.64 ± 4.18| | 1.30 ± 0.03 | 0.06 ± 0.01 |
IU: Image Uniformity, HCSR: High Contrast Spatial Resolution, IL: Image Linearity