. 2021 Apr 1;11(2):151–162. doi: 10.31661/jbpe.v0i0.2001-1061

Table 4.

Quantitative quality control (QC) measurements conducted on all considered devices.

Device Rater IU (GU) HCSR (10-1 mm) IL (%)
Kodak 210 P1 |17.79 ± 7.41| 0.83 ± 0.05 0.07 ± 0.01
P2 |14.62 ± 6.27| 0.83 ± 0.06 0.07 ± 0.01
D |15.38 ± 6.25| 0.90 ± 0.05 0.08 ± 0.01
Orix 70 P1 |18.07 ± 7.62| 0.85 ± 0.10 0.05 ± 0.01
P2 |18.52 ± 7.83| 0.70 ± 0.08 0.05 ± 0.01
D |16.35 ± 6.84| 0.86 ± 0.07 0.05 ± 0.01
Carestream CS 2100 P1 |5.58 ± 4.23| 1.40 ± 0.06 0.06 ± 0.01
P2 |4.94 ± 3.54| 1.43 ± 0.06 0.06 ± 0.01
D |5.24 ± 3.76| 1.30 ± 0.05 0.08 ± 0.01
Vatech EzRay Premium Vex S100W (1) P1 |6.79 ± 4.72| 1.25 ± 0.05 0.06 ± 0.01
P2 |8.86 ± 4.86| 1.25 ± 0.02 0.06 ± 0.01
D |11.10 ± 4.90| 1.23 ± 0.11 0.13 ± 0.01
Vatech EzRay Premium Vex S100W (2) P1 |7.42 ± 4.07| 1.20 ± 0.05 0.07 ± 0.01
P2 |8.96 ± 4.64| 1.30 ± 0.05 0.09 ± 0.01
D |8.64 ± 4.18| 1.30 ± 0.03 0.06 ± 0.01

IU: Image Uniformity, HCSR: High Contrast Spatial Resolution, IL: Image Linearity