Skip to main content
. 2021 Mar 25;11(4):593. doi: 10.3390/diagnostics11040593

Table 5.

Cut-off value with sensitivity and specificity for VNT titer 80 and 160.

Manufacturer Titer 80 Titer 160
Cut-Off Sensitivity Specificity AUC Cut-Off Sensitivity Specificity AUC
Diasorin 35.1 82% 100% 0.9341 73.2 64% 96% 0.8636
Beckman Coulter 2.39 49% 100% 0.8777 3.31 50% 96% 0.8435
Euroimmun 2.09 65% 100% 0.8806 3.06 55% 96% 0.8258
Abbott 4.26 35% 100% 0.8819 6.00 26% 96% 0.7168
Roche 68.4 59% 94% 0.8514 131.4 13% 96% 0.7111

Legend: The individual assays report the results as follows: Abbott as an index value (S/C); Diasorin as arbitrary units per milliliter (AU/mL); Roche as cut-off index (COI); Beckman Coulter as an index value (S/CO); Euroimmun as a ratio (number only); Cut-off is defined as the 95% percentile of the group below the titer 80 and 160; AUC, area under the curve.