Table 2.

Equivalent circuit parameters of skin-electrode interface.

Type of Electrodes Goodness of Fit Rd
(kΩ)
Re
(kΩ)
Ce
(nF)
RI
(kΩ)
CI
(nF)
Ri
(kΩ)
Ci
(nF)
Z1
(kΩ)
Z2
(kΩ)
Z3
(kΩ)
Z4
(kΩ)
Z
(kΩ)
1 0.99 1.92 252.00 41.90 1.92 N.A.6 16.80 117.00 1.92 152.00 1.92 16.30 172.00
2 0.98 2.72 124.00 25.80 79.30 6.22 1600.00 13,60 2.72 115.00 79.20 542.00 739.00
3 0.92 1.89 122.00 2.01 507.00 5.56 5,000,000.00 5.66 1.89 122.00 478.00 1410.00 2010.00
4 0.99 1.86 157.00 57.50 1.86 N.A.6 24.20 89.30 1.86 104.00 1.86 23.30 131.00
5 0.98 2.45 411.00 47.10 2.63 N.A.6 316.00 13.10 2.45 156.00 2.63 280.00 442.00

1 ① represents the standard Ag/AgCl electrode; 2 ② represents the electrode based on the top grain layer of cowhide; 3 ③ represents the electrode based on the split layer of cowhide; 4 ④ represents the electrode based on the top grain layer with lotion; 5 ⑤ represents the electrode based on the split layer with lotion; 6 There is no CI in the equivalent circuit of wet electrode, so CI is not applicable.