Table 1.

EV counts obtained for platelet-derived EVs by flow cytometry (Gallios vs. CytoFLEX LX) and by NTA (scatter mode vs. fluorescence mode after staining with CMO or LA-AF555).

Flow Cytometry
Device Events/µL LA+ EVs/µL LA+ EVs
[% of All Events in the EV Gate]
LA+CD41+ EVs/µL LA+CD41+ EVs
[% of All Events in the EV Gate]
Gallios 1.5 ± 0.5 × 106 1.4 ± 0.6 × 106 93 ± 4 1.3 ± 0.5 × 106 87 ± 6
CytoFLEX LX 6.0 ± 5.5 × 107 3.3 ± 2.5 × 107 64 ± 13 1.7 ± 0.9 × 107 40 ± 20
Nanoparticle Tracking Analysis
ZetaView PMX-110 Particles/µL
[Scatter Mode]
LA+ Particles/µL
[Fluorescent Mode]
LA+
[% of All Particles in Scatter]
CMO+ Particles/µL
[Fluorescent Mode]
CMO+
[% of All Particles in Scatter]
4.0 ± 1.7 × 108 3.3 ± 1.7 × 107 9.5 ± 4 1.8 ± 0.9 × 108 36 ± 7

n = 8; same EV batches used for all measurements.