Figure 5

Three examples of the spatial relationship between restenosis/thrombosis and strut fracture. The figure presents three cases of pattern 4 ((A,B): DES; (C): BRS) fracture associated with device failure ((A): thrombosis; (B,C): restenosis). The residual vessel lumen is presented in red.