Table 1. Comparison of Lumen Volume Measured by IVUS, SATM, and LATM.

IVUS SATM LATM P ICC
Lumen volume (mm3) 93.0 (70.0–116.3) 78.4 (57.2–99.6) 93.7 (73.0–114.5) 0.005
Pairwise comparisons* IVUS vs. SATM 0.015 0.872 (0.610–0.952)
IVUS vs. LATM 0.999 0.930 (0.840–0.971)
SATM vs. LATM < 0.001 0.916 (0.330–0.977)

Data are means. Data in parentheses are 95% confidence interval. *Bonferroni corrected. ICC = intraclass correlation coefficient, IVUS = intravascular ultrasonography, LATM = location-adaptive threshold method, SATM = scan-adaptive threshold method