Skip to main content
. 2020 Aug 26;2020(8):CD013705. doi: 10.1002/14651858.CD013705

Mertens 2020.

Study characteristics
Patient Sampling Single‐group study to estimate sensitivity and specificity for diagnosis of active disease:
‐ samples from patients suspected of SARS‐COV‐2 infections (n = 328)
Recruitment: random sampling of samples submitted to 3 laboratories
322/328 NP samples (NP swabs) were randomly selected
Prospective or retrospective: retrospectively
Number of samples (samples with confirmed SARS‐CoV‐2): 328 (132)
Patient characteristics and setting Setting: unclear; samples from university laboratories (discussion states that no outpatient population has been sampled, therefore assume inpatients and HCW samples)
Location: laboratories at Université Libre de Bruxelles (LHUB‐ULB), UZ Leuven and Centre Hospitalier Universitaire Sart‐Tilman (CHU) Liège
Country: Belgium
Dates: 19‐30 March 2020
Symptoms and severity: not reported
Demographics: not reported
Exposure history: unclear; 53/328 samples were from HCW
Index tests Test name: COVID‐19 Ag Respi‐Strip
Manufacturer: Coris BioConcept (Belgium)
Antigen target: SARS‐CoV and SARS‐CoV‐2 highly conserved nucleoprotein
Antibody: monoclonal antibodies directed against SARS‐CoV and SARS‐CoV‐2 highly conserved nucleoprotein antigen
Test method: immunochromatographic assay using colloidal gold (CGIA)
Samples used: remnant respiratory specimens (322 NP swabs, 4 NP aspirate and 2 BAL)
Transport media: NP: flocked swab + UTM 3 mL (or 1 mL of Amies) (Copan, Brescia, Italy);
NPA: 3 mL VTM (veal infusion broth (Difco, Becton Dickinson, Sparks, MD, USA) supplemented with bovine albumin (Sigma Aldrich, St Louis, MO, USA))
BAL: N/A
Sample storage: not described
Test operator: laboratory technician
Definition of test positivity: visible reddish‐purple band appearing at the Test line position (T)
Blinding reported: not stated
Timing of samples: not clear
Target condition and reference standard(s) Reference standard: qRT‐PCR: RealStar SARS‐CoV‐2 RT‐PCR Kit from Altona‐diagnostics with a cut‐off set at 40 Ct (LHUB‐ULB); Roche LC480 thermocycler using Taqman Fast Virus 1‐Step Master Mix (Thermo Fisher) (Liege); QuantStudio Dx (Thermo Fisher Scientific) or Panther Fusion (PF, Hologic, San Diego, USA) (UZ Leuven)
Definition of non‐COVID cases:
  • Genetic target(s): RealStar: not stated;

  • Taqman Fast Virus: RdRp and E genes

  • QuantStudio Dx; "slightly adapted" E‐gene

  • Panther Fusion: E gene and ORF1‐ab


Samples used: as for index test (respiratory specimens (322 NP swabs, 4 NP aspirate and 2 BAL)
Timing of reference standard: not stated; same samples as for index test but analysed at time of collection
Blinded to index test: yes (undertaken for diagnostic purposes at time of collection)
Incorporated index test: no
Flow and timing Time interval between index and reference tests: same samples used; discussion report 'some delay' between PCR and antigen testing
All participants received same reference standard: yes but different RT‐PCR kits
Missing data: no
Uninterpretable results: none reported; discussion reports some difficulties in visualising the strip through the closed tube requiring the lab technician to open the test tube in the laminar air flow cabinet and pull out the strip with forceps
Indeterminate results (index test): weak T lines considered positive
Indeterminate results (reference standard): none reported; sensitivity can be extracted for cases with Ct values < or > 25 (high vs lower viral load)
Unit of analysis: refers to participants
Comparative  
Notes Funding: not stated
Publication status: preprint (not peer‐reviewed)
Sourcepreprint server (medRxiv)
Author COI: the IVD medical device has been developed by the investigator Pascal Mertens, Henri Magein, and Justine Bouzet working for Coris BioConcept (potential conflict of interest declared even though they don’t have any share in this company); Thierry Leclipteux was involved in the development of this test and is the CEO of Coris
Bioconcept (potential conflict of interest declared). All scientific investigators that are external to Coris BioConcept declare having no conflict of interest.
Methodological quality
Item Authors' judgement Risk of bias Applicability concerns
DOMAIN 1: Patient Selection
Was a consecutive or random sample of patients enrolled? Yes    
Was a case‐control design avoided? Yes    
Did the study avoid inappropriate exclusions? Yes    
Did the study avoid inappropriate inclusions? Yes    
Could the selection of patients have introduced bias?   Low risk  
Are there concerns that the included patients and setting do not match the review question?     High
DOMAIN 2: Index Test (Antigen tests)
Were the index test results interpreted without knowledge of the results of the reference standard? Unclear    
If a threshold was used, was it pre‐specified? Yes    
Could the conduct or interpretation of the index test have introduced bias?   Unclear risk  
Are there concerns that the index test, its conduct, or interpretation differ from the review question?     High
DOMAIN 2: Index Test (Rapid PCR tests)
DOMAIN 3: Reference Standard
Is the reference standards likely to correctly classify the target condition? No    
Were the reference standard results interpreted without knowledge of the results of the index tests? Unclear    
Reference standard does not incorporate result of index test? Yes    
Could the reference standard, its conduct, or its interpretation have introduced bias?   High risk  
Are there concerns that the target condition as defined by the reference standard does not match the question?     High
DOMAIN 4: Flow and Timing
Was there an appropriate interval between index test and reference standard? Yes    
Did all patients receive the same reference standard? Yes    
Were all patients included in the analysis? Yes    
Did all participants receive a reference standard? Yes    
Were results presented per patient? Yes    
Could the patient flow have introduced bias?   Low risk