Skip to main content
. 2021 Apr 20;23(5):350–357. doi: 10.1089/dia.2020.0525

Table 1.

Accuracy of Bolus Delivery at Different Bolus Volumes (n = 225 for 0.2 and 1 U; n = 118 for 10 U)

Insulin pump 0.2 U
1 U
10 U
Mean deviation ± SD Individual boluses within ±15% of the target Mean deviation ± SD Individual boluses within ±15% of the target Mean deviation ± SD Individual boluses within ±15% of the target
ACS −3.3% ± 8.4% 88% +0.3% ± 5.5% 99% ±0.0% ± 0.8% 100%
A6 +3.3% ± 23.3% 40% +4.0% ± 16.0% 65% +3.1% ± 0.8% 100%
OP +1.5% ± 20.8% 57% ±0.0% ± 12.5%7 77%7 +0.3% ± 0.7%7 100%7

A6, A6 TouchCare® System; ACS, Accu-Chek® Solo micropump system; OP, Omnipod®; SD, standard deviation.