Table 1.
Accuracy of Bolus Delivery at Different Bolus Volumes (n = 225 for 0.2 and 1 U; n = 118 for 10 U)
| Insulin pump | 0.2 U |
1 U |
10 U |
|||
|---|---|---|---|---|---|---|
| Mean deviation ± SD | Individual boluses within ±15% of the target | Mean deviation ± SD | Individual boluses within ±15% of the target | Mean deviation ± SD | Individual boluses within ±15% of the target | |
| ACS | −3.3% ± 8.4% | 88% | +0.3% ± 5.5% | 99% | ±0.0% ± 0.8% | 100% |
| A6 | +3.3% ± 23.3% | 40% | +4.0% ± 16.0% | 65% | +3.1% ± 0.8% | 100% |
| OP | +1.5% ± 20.8% | 57% | ±0.0% ± 12.5%7 | 77%7 | +0.3% ± 0.7%7 | 100%7 |
A6, A6 TouchCare® System; ACS, Accu-Chek® Solo micropump system; OP, Omnipod®; SD, standard deviation.