Table 2.
Accuracy of Basal Rate Delivery at Different Basal Rates of Nine Measurements for 72 h Divided Into 1-h Windows
| Insulin pump | 0.1 U/h |
1 U/h |
||||
|---|---|---|---|---|---|---|
| Mean total deviation | SD (1 h windows) | 1-h windows within ±15% of the target | Mean total deviation | SD (1 h windows) | 1-h windows within ±15% of the target | |
| ACS | −5.3% | 25.0% | 51% | −1.9% | 5.4% | 98% |
| A6 | +4.8% | 28.4% | 48% | +3.4%a | 17.0% | 60% |
One measurement was stopped earlier due to an error notice.