Method Node Parameter Settings
MS Scan Properties Detector Type Orbitrap
Orbitrap Resolution 120,000
Mass Range Normal
Use Quadrupole Isolation True
Scan Range (m/z) 375-1500
RF Lens (%) 30
AGC Target custom
Normalized AGC Target 250%
Maximum Injection Time Mode custom
Maximum Injection Time (ms) 50
Microscans 1
Data Type Profile
Polarity Positive
Source Fragmentation False
MIPS Properties Monoisotopic Peak Determination Peptide
Intensity Properties Filter Type Intensity Threshold
Intensity Threshold 5.0e3
Charge State Properties Include charge state(s) 2–5
Include undetermined charge states False
Dynamic Exclusion Properties Exclude after n times 1
Exclusion duration (s) 60
Mass tolerance ppm
Low 10
High 10
Exclude isotopes True
Performed dependent scan on single charge state per precursor only False
Data-Dependent MS2 Scan Properties Isolation mode Quadrupole
Isolation Window (m/z) 1.4
Isolation Offset Off
Activation Type HCD
Collision Energy Mode Fixed
CID Collision Energy (%) 35
Detector Type Ion Trap
Ion Trap Scan rate Rapid
Mass Range Normal
Scan Range Mode Auto
AGC Target Standard
Maximum Injection Time Mode (ms) 35
Data Type Centroid