Chest compression (CC) parameters with and without TrueCPR device.

Parameter Manual CC CCs with TrueCPR P
CC rate [n × min−1] 129 [IQR 122–135] 112 [IQR 103–113] 0.002
CC depth [mm] 45 [IQR 43–48] 53 [IQR 51–55] < 0.001
Full release [%] 33 [IQR 29–38] 95 [IQR 76–99] < 0.001
Correct hand position [%] 94 [IQR 82–96] 95 [IQR 89–98] 0.185

IQR — interquartile range