. 2021 Apr 29;11:9295. doi: 10.1038/s41598-021-88853-w

Table 2.

Quantitative assessment of artifact reduction and surrounding liver parenchyma.

Corrected attenuation Corrected image noise
Hyperdense artifact Hypodense artifact Surrounding liver parenchyma
CI 337.9 ± 104.2 (−)479.2 ± 132.0 75.8 ± 25.5 7.0 ± 5.7
40 keV 1251.4 ± 380.4 (−)1074.6 ± 90.1 251.5 ± 101.6 34.1 ± 27.4
50 keV 772.7 ± 295.5 (−)928.5 ± 149.4 160.3 ± 62.8 19.4 ± 13.6
60 keV 508.9 ± 176.2 (−)711.3 ± 155.3 106.6 ± 40.7 18.4 ± 30.8
70 keV 330.1 ± 128.3 (−)516.4 ± 138.4 74.5 ± 28.5 8.1 ± 6.7
80 keV 230.4 ± 82.7 (−)378.2 ± 121.0 54.2 ± 22.2 5.6 ± 5.5
90 keV 160.4 ± 63.0 (−)284.1 ± 107.9 41.0 ± 19.1 4.5 ± 4.7
100 keV 113.0 ± 52.5 (−)218.8 ± 100.0 32.3 ± 17.6 4.0 ± 4.4
110 keV 79.2 ± 47.9 (−)173.2 ± 95.5 25.7 ± 17.3 3.8 ± 4.3
120 keV 54.9 ± 46.2 ( −)139.9 ± 93.5 21.1 ± 17.2 5.0 ± 7.9
130 keV 36.8 ± 46.0 ( −)115.2 ± 92.7 17.7 ± 17.3 3.8 ± 4.4
140 keV 23.0 ± 46.5 ( −)96.1 ± 92.4 15.1 ± 17.5 3.9 ± 4.5
150 keV 12.1 ± 47.3 ( −)73.2 ± 105.9 13.0 ± 17.7 3.9 ± 4.6
160 keV 3.6 ± 48.1 ( −)69.6 ± 92.7 11.5 ± 17.9 4.2 ± 4.4
170 keV (-)3.3 ± 48.9 ( −)60.1 ± 92.9 10.2 ± 18.0 4.1 ± 4.7
180 keV (-)9.0 ± 49.7 ( −)52.4 ± 93.2 9.1 ± 18.2 4.1 ± 4.7
190 keV (-)12.2 ± 49.5 ( −)44.6 ± 94.5 7.9 ± 20.5 4.1 ± 4.7
200 keV (-)17.5 ± 51.0 ( −)40.7 ± 93.7 7.5 ± 18.1 4.1 ± 4.8
P values
CI vs. 40–50 keV P < 0.001 P < 0.001 P < 0.001 P < 0.001
CI vs. 60 keV P < 0.001 P < 0.001 P > 0.05 P < 0.001
CI vs. 70 keV P > 0.05 P > 0.05 P > 0.05 P > 0.05
CI vs. 80 keV P > 0.05 P < 0.001 P > 0.05 P > 0.05
CI vs. 90–200 keV P < 0.001 P < 0.001 P < 0.001 P > 0.05

CI conventional images.

Data in italics indicate no statistical significance between any two energy levels.