. 2021 Apr 29;10(5):4. doi: 10.1167/tvst.10.5.4

Table 1.

Summary of Crosslinking Studies for Primary Myopic Corrections – Energy Parameters

Authors Arm Crosslinking Device Wavelength Total Fluence Irradiance Pulsed Illumination Treatment Time Treatment Zone (mm)
Kanellopoulos26 KXL II Device 365 nm 12.4 J/cm2 30 mW/cm2 N 4 min 4
Lim et al.19 Mosaic System 365 nm 10 or 15 J/cm2 45 mW/cm2 Y Nomogram based 4.5
Elling et al.28 Mosaic System 365 nm 10 or 15 J/cm2 30 mW/cm2 Y Nomogram based 4
Sachdev et al.31 Mosaic System 365 nm 15 J/cm2 30 mW/cm2 Y 16 min, 40 sec 4
Hout et al.26 Mosaic System 365 nm 15 J/cm2 30 mW/cm2 Y 16 min, 40 sec 6
Sachdev et al.29 Mosaic System 365 nm 15 J/cm2 30 mW/cm2 Y 16 min, 40 sec 4
Fredriksson et al.32 Arm 1 Mosaic System 365 nm 15 J/cm2 30 mW/cm2 Y 16 min, 40 sec 4
Arm 2 Mosaic System 365 nm 15 J/cm2 30 mW/cm2 Y 16 min, 40 sec 6
Arm 3 Mosaic System 365 nm 15 J/cm2 30 mW/cm2 Y 16 min, 40 sec 4
Naslund et al.33 Arm 1 Mosaic System 365 nm 15 J/cm2 30 mW/cm2 Y 16 min, 40 sec 4 mm homogenous zone
Arm 2 Mosaic System 365 nm 15 J/cm2 30 mW/cm2 Y 16 min, 40 sec 4 mm annular zone