Table 2.

Validity of measuring heart rate with Polar OH1 and Fitbit Charge 3 devices. Data are aggregated to a single data point for each of the activity domains. Data are analyzed for all data points (column “Overall”) and for each of the activity domains.

Device Overall (n=111a) Rest (n=19a) Cycling light (n=19a) Cycling hard (n=19a) Treadmill (n=19a) Sprint cycling (n=18a) Sprint running (n=17a)
Polar H10

Heart rate (beats·min-1), mean (SD) 106 (27) 62 (9) 89 (11) 119 (12) 124 (14) 128 (17) 120 (12)
Polar OH1

Heart rate (beats·min-1), mean (SD) 105 (27) 62 (9) 89 (10) 118 (12) 124 (14) 122 (17) 121 (15)

Mean bias (beats·min-1) 1 0 0 1 0 5 1

Limits of agreement (beats·min-1) −8 to 10 0 to 1 −3 to 2 −2 to 4 −1 to 2 −8 to 18 −16 to 15

MAPEb (%) 0.6 0.2 0.7 −0.8 0.2 −4.1 0.5

Correlation coefficient 0.954 0.983 0.974 0.985 0.992 0.807 0.67

95% CI of correlation coefficient 0.953 to 0.955 0.982 to 0.984 0.971 to 0.977 0.083 to 0.987 0.992 to 0.993 0.795 to 0.819 0.651 to 0.687
Fitbit Charge 3

Heart rate (beats·min-1), mean (SD) 97 (26) 61 (9) 81 (10) 100 (21) 123 (10) 109 (23) 112 (11)

Mean bias (beats·min-1) −9 −1 −7 −19 −2 −18 −8

Limits of agreement (beats·min-1) −41 to 23 −2 to 0 −36 to 22 −63 to 26 −13 to 10 −63 to 26 −28 to 12

MAPE (%) 7.37 −1.5 −7 −15 −1 −13.9 −6

Correlation coefficient 0.888 0.884 −0.056 0.183 0.924 0.771 0.496

95% CI of correlation coefficient 0.885 to 0.890 0.876 to 0.891 −0.113 to 0.002 0.127 to 0.238 0.921 to 0.927 0.756 to 0.784 0.471 to 0.520

an: number of data points analyzed for each domain.

bMAPE: mean absolute percentage error.