Table 2.
Validity of measuring heart rate with Polar OH1 and Fitbit Charge 3 devices. Data are aggregated to a single data point for each of the activity domains. Data are analyzed for all data points (column “Overall”) and for each of the activity domains.
| Device | Overall (n=111a) | Rest (n=19a) | Cycling light (n=19a) | Cycling hard (n=19a) | Treadmill (n=19a) | Sprint cycling (n=18a) | Sprint running (n=17a) | ||||||||
| Polar H10 | |||||||||||||||
|
|
Heart rate (beats·min-1), mean (SD) | 106 (27) | 62 (9) | 89 (11) | 119 (12) | 124 (14) | 128 (17) | 120 (12) | |||||||
| Polar OH1 | |||||||||||||||
|
|
Heart rate (beats·min-1), mean (SD) | 105 (27) | 62 (9) | 89 (10) | 118 (12) | 124 (14) | 122 (17) | 121 (15) | |||||||
|
|
Mean bias (beats·min-1) | 1 | 0 | 0 | 1 | 0 | 5 | 1 | |||||||
|
|
Limits of agreement (beats·min-1) | −8 to 10 | 0 to 1 | −3 to 2 | −2 to 4 | −1 to 2 | −8 to 18 | −16 to 15 | |||||||
|
|
MAPEb (%) | 0.6 | 0.2 | 0.7 | −0.8 | 0.2 | −4.1 | 0.5 | |||||||
|
|
Correlation coefficient | 0.954 | 0.983 | 0.974 | 0.985 | 0.992 | 0.807 | 0.67 | |||||||
|
|
95% CI of correlation coefficient | 0.953 to 0.955 | 0.982 to 0.984 | 0.971 to 0.977 | 0.083 to 0.987 | 0.992 to 0.993 | 0.795 to 0.819 | 0.651 to 0.687 | |||||||
| Fitbit Charge 3 | |||||||||||||||
|
|
Heart rate (beats·min-1), mean (SD) | 97 (26) | 61 (9) | 81 (10) | 100 (21) | 123 (10) | 109 (23) | 112 (11) | |||||||
|
|
Mean bias (beats·min-1) | −9 | −1 | −7 | −19 | −2 | −18 | −8 | |||||||
|
|
Limits of agreement (beats·min-1) | −41 to 23 | −2 to 0 | −36 to 22 | −63 to 26 | −13 to 10 | −63 to 26 | −28 to 12 | |||||||
|
|
MAPE (%) | 7.37 | −1.5 | −7 | −15 | −1 | −13.9 | −6 | |||||||
|
|
Correlation coefficient | 0.888 | 0.884 | −0.056 | 0.183 | 0.924 | 0.771 | 0.496 | |||||||
|
|
95% CI of correlation coefficient | 0.885 to 0.890 | 0.876 to 0.891 | −0.113 to 0.002 | 0.127 to 0.238 | 0.921 to 0.927 | 0.756 to 0.784 | 0.471 to 0.520 | |||||||
an: number of data points analyzed for each domain.
bMAPE: mean absolute percentage error.