Skip to main content
. 2021 Mar 25;9(3):e25313. doi: 10.2196/25313

Table 3.

The mean absolute percentage error for each participant for Polar OH1 and Fitbit Charge 3 across all available data points.

Participant Polar OH1 (%) Fitbit Charge 3 (%)
1 0.22 6.34a
2 −3.92 5.30a
3 1.37 2.80
4 0.06 2.06
5 0.31 4.01
6 1.70 1.53
7 0.22 2.35
8 −0.44 2.59
9 0.71 3.14
10 −0.09 2.83
11 0.59 -3.25
12 1.81 4.00
13 −0.10 3.91
14 −0.09 −5.43a
15 0.69 14.90a
16 0.87 5.97a
17 0.19 2.75
18 1.18 11.40a
19 2.58 15.80a
Overall 0.41 4.37

aExceeds 5% mean absolute percentage error threshold.