TABLE 4.

Performance improvement of anti-HTLV-1 Ab assays for CSF by the best cutoff point

Performance characteristic CLIA performance at indicated cutoff point
CLEIA performance at indicated cutoff point
S/CO of 1.00 (default, adopted from serum) S/CO of 1.50 (besta) COI of 1.00 (default, adopted from serum) COI of 0.14 (besta)
Sensitivity (%) 96.6 94.8 69.0 89.7
Specificity (%) 86.4 95.5 100.0 95.5
Accuracy (%)b 93.8 95.0 77.5 91.3
Likelihood ratioc 7.1 21.1 NAd 19.9
a

The best cutoff points for anti-HTLV-1 Ab assay in CSF were determined by ROC curve analysis (Fig. 3B).

b

Accuracy was calculated as (true positive + true negative)/(total number).

c

Likelihood ratio was defined as sensitivity/(1 − specificity).

d

NA, not applicable because the specificity of CLEIA at the default cutoff point was 100%.